Hall D Maintenance Status Long Description

Click the Last Maintenance date to update when the last maintenance was performed

Click the Task to EDIT the long description of this task

Click here to add a new item to the Hall D maintenance maintainer

TASK Primary Contact Secondary Contact PERIOD (days) NEXT MAINTENANCE LAST MAINTENANCE
Vacuum Pump #004 keithb@jlab.org bunton@jlab.org 365 11/29/24 11/30/23

Pfeiffer Model DUO 10M F224161 Serial# 22016984 Check oil level, change if dirty. Inspect and clean exterior of pump. Operate for 1 hour and record ultimate vacuum pressure. Pressure on 02-03-16 after 15mins= 2.01 mTorr. Spare pump ,located in Hall D tagger at east wall turbo cart.