Hall D Maintenance Status Long Description

Click the Last Maintenance date to update when the last maintenance was performed

Click the Task to EDIT the long description of this task

Click here to add a new item to the Hall D maintenance maintainer

TASK Primary Contact Secondary Contact PERIOD (days) NEXT MAINTENANCE LAST MAINTENANCE
Vacuum Pump #005 keithb@jlab.org bunton@jlab.org 365 11/29/24 11/30/23

Pfeiffer Model DUO 10M F224124 Serial# 22018702 Check oil level, change if dirty. Inspect and clean exterior of pump. Run for 1 hour and record ultimate vacuum pressure. Pressure on 10/20/16 after 15mins=3.1 mtorr Pump located below F-cal platform. 2/7/2019 oil ok 3/5/2020 oil good 8-3-2021 - Oil changed