(Created page with "__TOC__ == Set up harp scan == # Ask MCC to stop beam delivery in preparation for harp scans and to turn off tagger quadrupole (MQPAD00) # Remove all radiators # Move col...") |
(→Restore Production conditions) |
||
(One intermediate revision by the same user not shown) | |||
Line 30: | Line 30: | ||
# Perform radiator harp scan as described in [https://halldweb.jlab.org/hdops/wiki/index.php/Beam_Line_Detectors_Shift#Radiator_Harp_Scan this wiki page] | # Perform radiator harp scan as described in [https://halldweb.jlab.org/hdops/wiki/index.php/Beam_Line_Detectors_Shift#Radiator_Harp_Scan this wiki page] | ||
− | == Restore Production conditions | + | == Restore Production conditions == |
# Ask MCC to take beam away and mask FSD for radiator movement | # Ask MCC to take beam away and mask FSD for radiator movement | ||
− | # Insert | + | # Insert desired radiator (diamond or 3x10^-4 RL amorphous) |
# Move collimator to 5.0 mm diameter hole (can be done in parallel with previous step) | # Move collimator to 5.0 mm diameter hole (can be done in parallel with previous step) | ||
# Ask for the appropriate current CW beam to be restored | # Ask for the appropriate current CW beam to be restored | ||
# Refer to [https://halldweb.jlab.org/hdops/wiki/index.php/Standard_Production_Running_Conditions this wiki page] for standard production conditions | # Refer to [https://halldweb.jlab.org/hdops/wiki/index.php/Standard_Production_Running_Conditions this wiki page] for standard production conditions |
Latest revision as of 18:01, 7 March 2017
Contents
Set up harp scan
- Ask MCC to stop beam delivery in preparation for harp scans and to turn off tagger quadrupole (MQPAD00)
- Remove all radiators
- Move collimator to blocking position
Perform accelerator harp scans (quad off)
- Ask MCC for three harp scans: 5C11, 5C11B, and tagger dump (AD00)
- Remind MCC that all three should be done with tune beam
Perform radiator harp scan (quad off)
- Ask for 100 nA CW beam
- Perform radiator harp scan as described in this wiki page
Perform accelerator harp scans (quad on)
- Ask MCC to turn tagger quadrupole (MQPAD00) back on with current of -4.2A
- Ask MCC for three harp scans: 5C11, 5C11B, and tagger dump (AD00)
- Remind MCC that all three should be done with tune beam
Perform radiator harp scan (quad on)
- Ask for 100 nA CW beam
- Perform radiator harp scan as described in this wiki page
Restore Production conditions
- Ask MCC to take beam away and mask FSD for radiator movement
- Insert desired radiator (diamond or 3x10^-4 RL amorphous)
- Move collimator to 5.0 mm diameter hole (can be done in parallel with previous step)
- Ask for the appropriate current CW beam to be restored
- Refer to this wiki page for standard production conditions