Difference between revisions of "CDC and FDC readout tests"
From GlueXWiki
(Created page with "The goal is to take CDC data using the FDC amplitude mode (FDC_amp_short) which is data type 9, I’ll have to check which mode # includes that. The reason for this is that...") |
|||
(One intermediate revision by the same user not shown) | |||
Line 1: | Line 1: | ||
− | The goal is to take CDC data using the FDC amplitude mode (FDC_amp_short) which is data type 9, | + | The goal is to take CDC data using the FDC amplitude mode (FDC_amp_short) which is data type 9, included in mode 5. |
The reason for this is that we switched from using pulse integral to amplitude for the CDC dE/dx, and using the FDC_amp_short mode will give us an extra 3 bits to read out amplitude and an extra 2 for pedestal, but not read out the integral. | The reason for this is that we switched from using pulse integral to amplitude for the CDC dE/dx, and using the FDC_amp_short mode will give us an extra 3 bits to read out amplitude and an extra 2 for pedestal, but not read out the integral. | ||
Line 6: | Line 6: | ||
*With beam, at the beginning of the run period: | *With beam, at the beginning of the run period: | ||
− | *10 evio files (max 1h) in FDC_amp_short with abit=pbit=0 (no scaling down) | + | *10 evio files (max 1h) in FDC_amp_short (mode 5) with abit=pbit=0 (no scaling down) |
− | *10 evio files (max 1h) in CDC_short | + | *10 evio files (max 1h) in CDC_short (mode 3) with abit=3 pbit=0 |
*Runs must be done in the same conditions (weather & beam intensity). 1h per mode is assumed, but if things go well it will only take 10 mins . | *Runs must be done in the same conditions (weather & beam intensity). 1h per mode is assumed, but if things go well it will only take 10 mins . | ||
− | *Then for the rest of the run we should use FDC_amp_short. | + | *Then for the rest of the run we should use FDC_amp_short (mode 5). |
Naomi will look at the test data (the 1h of each mode as above). If there is any problem with it, we can switch back to CDC_short. | Naomi will look at the test data (the 1h of each mode as above). If there is any problem with it, we can switch back to CDC_short. |
Latest revision as of 08:54, 30 November 2017
The goal is to take CDC data using the FDC amplitude mode (FDC_amp_short) which is data type 9, included in mode 5.
The reason for this is that we switched from using pulse integral to amplitude for the CDC dE/dx, and using the FDC_amp_short mode will give us an extra 3 bits to read out amplitude and an extra 2 for pedestal, but not read out the integral.
Plan:
- With beam, at the beginning of the run period:
- 10 evio files (max 1h) in FDC_amp_short (mode 5) with abit=pbit=0 (no scaling down)
- 10 evio files (max 1h) in CDC_short (mode 3) with abit=3 pbit=0
- Runs must be done in the same conditions (weather & beam intensity). 1h per mode is assumed, but if things go well it will only take 10 mins .
- Then for the rest of the run we should use FDC_amp_short (mode 5).
Naomi will look at the test data (the 1h of each mode as above). If there is any problem with it, we can switch back to CDC_short.