Difference between revisions of "Newest preamp"
From GlueXWiki
Line 5: | Line 5: | ||
*2050V | *2050V | ||
*Comparison of modified preamp (preamp4 with terminating resistors removed) and its predecessor (preamp4) | *Comparison of modified preamp (preamp4 with terminating resistors removed) and its predecessor (preamp4) | ||
+ | |||
Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below) Blue histograms are tracked hits, runtimes were not equal. | Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below) Blue histograms are tracked hits, runtimes were not equal. | ||
Line 12: | Line 13: | ||
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" | ||
|width="420pt"| | |width="420pt"| | ||
− | [[Image:run_31599_amp6.png|thumb|420px|Max amplitude histo for cosmics, new fADC, modified HVB, mixed 50/50 at 60 sccm 2050V]] |} | + | [[Image:run_31599_amp6.png|thumb|420px|Max amplitude histo for cosmics, new fADC, modified HVB, mixed 50/50 at 60 sccm 2050V]] |
+ | |} | ||
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" |
Revision as of 08:44, 20 April 2011
- 50/50 mixed Ar, CO2
- Recalibrated MFCs
- Outer plenum added to CDC prototype to prevent leaks
- New modified HVB
- 2050V
- Comparison of modified preamp (preamp4 with terminating resistors removed) and its predecessor (preamp4)
Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below) Blue histograms are tracked hits, runtimes were not equal.
The fADC range is 0-4096. These are histograms of the first maximum in ADC value for each event, pedestal has been subtracted.
Pedestal mean was approx 100, sigma is 9-12.
Earlier new fADC results are on this page: New_fADC_with_cosmics
Below: 55Fe spectra with preamp 4 and Struck ADCs set to half-range, 550mV
Below: 55Fe spectra with newest preamp (preamp4 minus terminating resistors) and Struck ADCs set to full-range, 1.1V