Difference between revisions of "Pre-production prototype HVB"
From GlueXWiki
Line 6: | Line 6: | ||
*Comparison of pre-production modified CDC HVB prototype from manufacturer with modified HVB from Fernando (the two should be identical) | *Comparison of pre-production modified CDC HVB prototype from manufacturer with modified HVB from Fernando (the two should be identical) | ||
− | Below: max amplitude histograms (left) and | + | Below: max amplitude histograms (left) and amplitude integrated over 96ns (right) from 55Fe, new prototype HVB, cosmics rejected in software |
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" | ||
|width="420pt"| | |width="420pt"| | ||
Line 13: | Line 13: | ||
[[Image:run_31677_e.png|thumb|420px|Integrated amplitude histo for 55Fe, pre-production prototype HVB]] | [[Image:run_31677_e.png|thumb|420px|Integrated amplitude histo for 55Fe, pre-production prototype HVB]] | ||
|} | |} | ||
− | Below: max amplitude histograms (left) and | + | Below: max amplitude histograms (left) and amplitude integrated over 96ns (right) from 55Fe, Fernando's modified HVB, cosmics rejected in software |
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" | ||
|width="420pt"| | |width="420pt"| |
Revision as of 14:04, 17 June 2011
- 50/50 mixed Ar, CO2 at 60 sccm
- Recalibrated MFCs
- Outer plenum added to CDC prototype to prevent leaks
- 2050V
- Modified preamp (preamp4 with terminating resistors removed)
- Comparison of pre-production modified CDC HVB prototype from manufacturer with modified HVB from Fernando (the two should be identical)
Below: max amplitude histograms (left) and amplitude integrated over 96ns (right) from 55Fe, new prototype HVB, cosmics rejected in software
Below: max amplitude histograms (left) and amplitude integrated over 96ns (right) from 55Fe, Fernando's modified HVB, cosmics rejected in software