Revision as of 19:11, 17 March 2016
Run | hits | repeatfree hits | diffs | time | q | pedestal | amplitude | integral | overflow count | readout | |
10954 | 28658293 | 24179315 | 12394 | 61 | 0 | 1 | 12326 | 6 | 0 | CDC | |
10956 | 87414131 | 73914777 | 145176 | 107559 | 275 | 4 | 37530 | 12672 | 0 | CDC | |
10957 | 19196556 | 16340197 | 33186 | 24069 | 59 | 3 | 9086 | 2868 | 0 | CDC & FDC |
NB hits with repeated pairs are not necessarily an error, it was just simpler to exclude all of them
Most of the discrepancies for run 10965 come from 6 channels where the firmware time = 1 + emulated time, they are not noise events.
Run | hits | repeatfree hits | diffs | time | q | pedestal | amplitude | integral | overflow count | readout | |
10986 | 8058453 | 6875279 | 167 | 15 | 0 | 0 | 150 | 2 | 0 | CDC & FDC |