FA125 firmware check
Used the raw samples to emulated the fa125's calculated values and compared them with the fa125 output.
Only looked at the first evio file so far.
Run | events | time | amplitude | pedestal | integral | overflow count |
3293 | 16582555 | 1160 | 1942 | 3547 | 15609 | 1 |
3293 minus 2 bad fadcs | 15914850 | 378 | 1877 | 288 | 12798 | 1 |
4623 | 10247801 | 0 | 98 | 0 | 3 | 0 |
Many of the problems in 3293 were due to hardware faults in roc28 slots 5&6.
12205 of the 15609 difference in integral were due to faulty assignment of the overflow bit. This has been fixed.
There were a few problems in the firmware which Cody described last week. Almost all fixed.
Pulse & Window raw data are separate objects in the evio eventloop, not linked yet. One out of sync pair causes the rest of the data for that trigger to be out of step.
Run 3923 had 7600 entries with pulse and WRD out of step due to 252 hiccups.
Run 4623 had 1317 entries with pulse and WRD out of step due to 33 hiccups.