Active Base Radiation Tests

From GlueXWiki
Revision as of 13:41, 30 November 2023 by Aoranliu (Talk | contribs) (Active Divider Radiation Test)

Jump to: navigation, search

Active Divider Radiation Test

Measurements
Date Total Dose (kR) Divider current1 (uA) Pulse Amplitude1 (mV) Comment1 Divider current2 (uA) Pulse Amplitude2 (mV) Comment2
2023/11/08 0 kR 388.5 998 mV Positive Voltage 4V, Negative Voltage -4V (in contrast)[1] 388.5 998 mV Positive Voltage 4V, Negative Voltage -4V (in contrast)[2]
2023/11/08 3.2 kR 386.5 951 mV 386.5 937 mV
2023/11/09 7 kR 384.5 923 mV 384.0 898 mV
2023/11/10 10.22 kR 382.5 877 mV 382.5 860 mV
2023/11/13 12.4 kR 381.5 858 mV 380.5 845 mV
2023/11/14 15.6 kR 379.5 825 mV 378.0 787 mV
2023/11/15 19 kR 377.5 142 mV Positive Voltage 4.2V, Negative Voltage -1.7V [3] 376.5 1064 mV Positive Voltage -0.7V, Negative Voltage -3.4V, weird tail in waveform[4]
378.5 808 mV Regulator removed [5] 377.0 765 mV Regulator removed [6]
2023/11/16 22.3 kR 376.5 771 mV Waveform [7] 375.0 740 mV Waveform[8]
2023/11/17 26.2 kR 375.0 744 mV Waveform [9] 373.0 718 mV Waveform[10]
2023/11/20 28.4 kR 375.0 748 mV Waveform [11] 373.0 712 mV Waveform[12]
2023/11/21 31.9 kR 373.5 727 mV Waveform [13] 371.0 686 mV Waveform[14]
2023/11/22 35.25 kR 372.5 715 mV Waveform [15] 370.5 682 mV Waveform[16]
2023/11/29 38.66 kR 372.5 708 mV Waveform [17] 370.5 668 mV Waveform[18]
2023/11/30 42.1 kR 371.0 682 mV Waveform [19] 369.0 643 mV Waveform[20]

Spreadsheet from RadCon