CDC gain calibrations

From GlueXWiki
Revision as of 15:53, 24 June 2016 by Njarvis (Talk | contribs)

Jump to: navigation, search

Early runs 10391-10913

  • fadc thresholds were high and for runs with high atmospheric pressure the minimum ionizing peak in pulse amplitude was partly or mostly below threshold
  • Chose straws 501-510 as these had high gain and their MIP was visible in most runs, found the scaling factor to match their MPV for the run of interest to run 11059, scaled the individual gains for run 11059 to give gains for the run of interest. Chose run 11059 as it was the next run after the thresholds had been adjusted.
Pressure during runs 10331-10913. Circled runs selected for calibration.
Temperature during runs 10391-10913
run11059 amplitude
run10391 amplitude
run10518 amplitude
run10562 amplitude
run10564 amplitude
run10780 amplitude
run10837 amplitude
MIP MPV vs pressure
MIP MPV vs pressure/temperature


Disconnected straws: 709, 2384, runs 10331 on.


Other missing straws:

straw 3150 runs 10455-11059
straw 3151 runs 10331-11059

straws 21-25, 62-67, 110-116, 164-169 (HVB A5) runs 10392-10399

straws 2392-2399 runs 10455-10464  (HVB I8 and another)
straws 2575-2582 runs 10455-10464
straws 2760-2767 runs 10455-10464
straws 2957-2964 runs 10455-10464
straws 3157-3164 runs 10455-10464
straws 3366-3373 runs 10455-10464


straw 2384-2391 run 10518 (not approved, but interesting for the low pressure)
straw 2567-2574 run 10518
straw 2751-2759 run 10518
straw 2950-2956 run 10518
straw 3149-3156 run 10518
straw 3358-3365 run 10518


one straw from row 23 missing in 10573


straw 2873 missing in 011059
straw 2964 missing in 011059
straw 3372 missing in 011059


Low gain straws:

straw 244  runs 10331-11059 
straw 1391 runs 10391-
straw 2977  runs 10331-11059 
straw 3178  runs 10331-11059 


straw 3117 no amplitudes above 488 run 010562
straw 3155 no amplitudes above 490 run 010562
straw 3240 no amplitudes above 490 run 010562
straw 3248 no amplitudes above 486 run 010562


Low-gain HVB group 3064-3070 
Low-gain HVB group 3270-3277
Low-gain HVB group 3479-3486




run 10703 - occupancy plot suggests that HVB A6 has low efficiency, the MIP is below threshold but the preamp saturation peak is not visible, so the preamp is probably ok but the gain is low. Check the HV. A6 includes straws 21-25.

[ https://halldweb.jlab.org/work/halld2/data_monitoring/RunPeriod-2016-02/mon_ver03/Run010703/CDC_occupancy.png occupancy plot ]

run 10703 straw 20
run 10703 straw 22