Difference between revisions of "CDC more garfield calcs"

From GlueXWiki
Jump to: navigation, search
Line 16: Line 16:
 
{| border="0" cellpadding="2"  
 
{| border="0" cellpadding="2"  
 
|+ Real life results from prototype 2100V 50/50 Ar/CO2 with scintillators vertical and track angle > 85^{o}
 
|+ Real life results from prototype 2100V 50/50 Ar/CO2 with scintillators vertical and track angle > 85^{o}
|[[Image:cdc_run_31962_amp_85deg.png|thumb|x200px|First max in signal amplitude]]
+
|[[Image:cdc_run_31962_amp_85deg.png|thumb|x200px|First max in signal amplitude, track angle > 85^{o}]]
|[[Image:cdc_run_31962_int_85deg.png|thumb|x200px|Integrated signal]]
+
|[[Image:cdc_run_31962_int_85deg.png|thumb|x200px|Integrated signal, track angle > 85^{o}]]
 
|}
 
|}

Revision as of 15:51, 18 April 2014

Cosmics from prototype data of run 31962, 50/50 Ar/CO2, 2100V, scintillators above and below prototype with paddles vertical Picked events with hits in 3 straws in same vertical column in prototype. Estimate from detector geometry that track angle < 14 degrees to vertical.

Real life results from prototype 2100V 50/50 Ar/CO2 with scintillators vertical
First max in signal amplitude
Integrated signal
Garfield, 3GeV u at 0-14degrees, gain 3.84e6 (gas gain) x 0.058 (transfer function)
First max in signal amplitude
Integrated signal
Real life results from prototype 2100V 50/50 Ar/CO2 with scintillators vertical and track angle > 85^{o}
First max in signal amplitude, track angle > 85^{o}
Integrated signal, track angle > 85^{o}