Difference between revisions of "CDC preamp LV checkup"

From GlueXWiki
Jump to: navigation, search
Line 8: Line 8:
 
|}
 
|}
 
#pedestal noise and stability
 
#pedestal noise and stability
 +
{| border="0" cellpadding="2"
 +
|[[Image:run_31784_pedmean.png|thumb|x150px|Pedestal mean of samples 0-99, ch7, over 2.5days ]]
 +
|[[Image:run_31784_pedsig.png|thumb|x150px|Pedestal std dev, ch 7 ]]
 +
|}
 
#tails from large pulses
 
#tails from large pulses
 
#gain
 
#gain
 
#efficiency
 
#efficiency
 
#resolution
 
#resolution

Revision as of 11:34, 11 May 2012

Following a tip-off from Lubomir that the LV reaching the preamp card might not be correct, I checked it at the board and found it to be 2.6V rather than 3V. This is due partly to voltage drop along the cable and partly to the preamp taking more current than the supply can put out. The original LV supply unit was a HPsomething. I switched it first for a Lambda LH121FM which could put out enough current (0.5A) at 3V but found this to have a wavering pedestal - +/-30 adc bins over the course of 24h, and then for a Global Specialties 1305 which is looking good after 3d. Both have more pedestal noise (pedestal s.d. ~15 bins) than the HP (~10 bins).

Things to check:

  1. pedestal height - this was previously set to ch 100 - adjust offsets to bring this back to 100
ADC contents, one event (noise), first 3 are logic signals, rest are CDC prototype
  1. pedestal noise and stability
Pedestal mean of samples 0-99, ch7, over 2.5days
Pedestal std dev, ch 7
  1. tails from large pulses
  2. gain
  3. efficiency
  4. resolution