Difference between revisions of "CDC tilted detectors"
Line 212: | Line 212: | ||
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" | ||
− | |+ Prototype tilted | + | |+ Prototype tilted 0deg |
|[[Image:run_31989_ampfit.png|thumb|x150px|Max amplitude]] | |[[Image:run_31989_ampfit.png|thumb|x150px|Max amplitude]] | ||
|[[Image:run_31989_integfit.png|thumb|x150px|Integration]] | |[[Image:run_31989_integfit.png|thumb|x150px|Integration]] | ||
Line 230: | Line 230: | ||
The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution. | The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution. | ||
− | |||
− | |||
Line 294: | Line 292: | ||
| 1.16 ± 0.02 | | 1.16 ± 0.02 | ||
| 1.32 | | 1.32 | ||
+ | |} | ||
+ | |||
+ | |||
+ | |||
+ | {| border="0" cellpadding="2" | ||
+ | |+ Integrated signal histograms, Drift time < 11ns | ||
+ | |[[Image:run_31989_integfit_11ns.png|thumb|x150px|Tilt 0°]] | ||
+ | |[[Image:run_31986_integfit_11ns.png|thumb|x150px|Tilt 30°]] | ||
+ | |[[Image:run_31984_integfit_11ns.png|thumb|x150px|Tilt 45°]] | ||
|} | |} |
Revision as of 18:50, 14 February 2013
Prototype at 2100V, 50/50 Ar/CO2
The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution.
Ratio of MPVs of Landau fit to integration histograms for horizontal : 30° : 45° should be 1 : 1.15 : 1.41
The 45° and 0° runs were separated by 1/2 day; the 30° data was taken one week later.
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
218 | 229 | 256 | 264 | 264 | 285 | 1.17 | 1.15 | 1.21 | 1.24 |
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
5167 | 5301 | 6722 | 7446 | 7163 | 8113 | 1.30 | 1.40 | 1.39 | 1.53 |
The events in the noise peak in the amplitude histo do not reach the integration histo because the code ignores hits with a leading edge gradient less than 1.
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
281 | 272 | 364 | 335 | 382 | 383 | 1.30 | 1.23 | 1.36 | 1.41 |
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
8604 | 7986 | 11658 | 10839 | 12339 | 11537 | 1.35 ± 0.04 | 1.36 | 1.43 ± 0.04 | 1.44 |
50/50 premix gas, unfiltered, label reads analysis 49.1 CO2, so this is 51/49 Ar/CO2
2100V with prototype horizontal, scintillators vertical
Estimated max drift time is 650ns
Repeating the tilted measurements just in case the scintillators moved slightly between measurements
The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution.
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
383 | 403 | 394 | 419 | 378 | 413 | 1.03 | 1.04 | 0.99 | 1.02 |
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
8038 | 7994 | 9049 | 10413 | 9291 | 10591 | 1.13 ± 0.02 | 1.30 | 1.16 ± 0.02 | 1.32 |