Calibration Dashboard

From GlueXWiki
Revision as of 22:33, 10 May 2015 by Sdobbs (Talk | contribs)

Jump to: navigation, search

Fall 2014 Data Calibration Status

This table summarizes the status of the calibrations that have been stored in the CCDB.

Last Updated -- 17 Feb. 2015

  • red - This category has not been calibrated
  • yellow - Calibrations have been partially done, or done with cosmic ray or other non-photon beam data
  • green - Calibrations have been done with photon beam data


Detector Base Time fADC E scale fADC gains fADC timing TDC timing TDC timewalk Reconstructed Objects Alignment
BCAL N/A 1 2 1
CDC N/A N/A 1
FCAL 1 1 N/A N/A
FDC N/A N/A
PS/PSC
Start Counter 1
TAGH 1 1
TAGM 1 1
TOF 1

Notes:

  • "Base time" refers to the relative timing between different detectors.
  • The fADC and TDC categories refer to per-channel calibrations.
  • The high-level reconstruction objects are the final inputs to a physics analysis. The low-level reconstruction objects are whatever objects are built from the calibrated hits and used to build the high-level objects.


Performance

Values obtained from data are shown in blue.

Detector Timing resolution Energy Resolution Spatial Resolution Efficiency
BCAL 1.8 ns / 70 ps/sqrt(E) ~19 MeV (pi0 mass, E(pi0)>1.5 GeV)/ 5.4%/sqrt(E) + 2.3%, 8.7 MeV (pi0 mass) 6 cm (ADC), 35 cm (TDC) [z] / ~8 mrad (trans.); 1.1 cm/sqrt(E) [z] > 90%
CDC / / ~150/250 um (far/close to wire) / ~200 um <~95%
FCAL 3.2 ns / ~0.4 ns ~11 MeV (pi0 mass) / 5.6%/sqrt(E) + 3.5%, 8.7 MeV (pi0 mass) / 6.4 mm/sqrt(E)
FDC / / 150-250 um / ~200 um ~95%
PS/PSC / / /
Start Counter ~480 ps / 350 ps / / ~90%
TAGH 1.6 ns / / / ~85%
TAGM 1.6 ns / / / <~70%
TOF ~150 ps / ~90 ps / ~1.2 cm /

Notes:

  • When possible, the results are given as "Measured / Design"