Construction of Low-Granularity counters

From GlueXWiki
Revision as of 15:39, 24 June 2013 by Dzbenski (Talk | contribs) (Created page with "==PMT Testing== ===Expectations=== The photomultiplier tubes (i.e. 18 in all) should, in principle, all exhibit the same characteristics. All 18 should have the same change in ou...")

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

PMT Testing

Expectations

The photomultiplier tubes (i.e. 18 in all) should, in principle, all exhibit the same characteristics. All 18 should have the same change in output amplitude with a change in input potential, as well as identical time responses regardless of the input potential. Therefore, one would expect a histogram of these characteristics to show a count of 18 at a single value on the x-axis. Consequently, a logarithmic plot of the overall gains of each PMT should be 18 identical lines of positive slope.

The Hamamatsu Photomultiplier Tube (H7415) was chosen because it has the characteristics required for this pair spectrometer. This PMT will be attached to the EJ-200 organic plastic scintillator and light guide fabricated by Eljen Technology. The properties of the tube are in the table below:


Photomultiplier Tube Assembly (H7415)
Dynode StructureLinear-focused
Dynode Stages10
Max. Rating) Anode to Cathode Voltage-2000 V
Anode to Cathode Supply Voltage-1500 V
Anode) Gain Typ.5E+06
(Time Response) Rise Time Typ.1.7 ns
(Time Response) Transit Time Typ.16 ns


Rise time listed here was arrived at using a laser. In the testing of the PMT's at Jefferson Lab, an LED will be used, which will not produce a clean signal elimination like the laser.

Testing

To test the PMT's, connections had to be soldered on to the wires in order to have quick connection ability to both the oscilloscope and high-voltage power source. Once this was done, the PMT was placed in a previously fabricated dark box with mounts for the tube and LED. The LED was powered by a wave generator which was set to run 8V square waves at 10 kHz with a 8 ns pulse width. Figure 1 shows the set up used for each PMT.

Figure 1. Testing configuration for PMTs.

The LED was positioned behind a diffracted lens to spread the light out. Careful attention had to be made that each PMT was positioned at the same place on the mount in order to eliminate fluctuations in beam intensity. The anode-to-cathode potential would range from -1.1 kV to -1.5 kV.

Data

The data gathered from the oscilloscope were the amplitude and rise time. Figure 2 shows what those values represent on the oscilloscope screen. Amplitudes were measured in millivolts and the rise time was in nanoseconds.

Figure 2. Oscilloscope analysis was done using the rise time and amplitude.

The data for the PMT amplitudes is in the table below. At -1.1 kV, the minimum amplitude was 124 mV. At -1.5 kV, the maximum amplitude was 1660 mV.


PMT Amplitudes (mV)
PMT #-1.1 kV-1.2 kV-1.3 kV-1.4 kV-1.5 kV
11713035138331320
2124219365583906
31873305538921400
421337662710101575
521638765710601670
61783135228391310
71713025078121256
81813225428771380
91592844737551160
101682995048111260
111602894897851236
1221037062410001556
1321438565010501660
142033596009681500
151592874837801215
161702995048101300
171953486079701500
182163766279971550