Difference between revisions of "Discussion of e+/e- tertiary beams in Hall D"

From GlueXWiki
Jump to: navigation, search
(Running conditions of 2-photon experiment in Hall B)
(Running conditions of 2-photon experiment in Hall B)
Line 18: Line 18:
 
*: max power in photon beam: 12 W
 
*: max power in photon beam: 12 W
  
This is only a factor 2.5 higher absorbed dose in the collimator cave than expected for high-luminosity running in GlueX.  A more conservative configuration with the same photon beam power would be the following.
+
The radiator thickness was increased just enough to get the same photon beam intensity as the Hall B experiment.  Notice that the total photon beam is only a factor 2.5 higher absorbed dose in the collimator cave than expected for high-luminosity running in GlueX.  A more conservative configuration with the same photon beam power might be the following, with the same photon beam intensity.
 
*: primary beam energy: 12 GeV
 
*: primary beam energy: 12 GeV
 
*: max beam current: 300 nA
 
*: max beam current: 300 nA
Line 31: Line 31:
 
*: radiator thickness: 0.0025 radiation lengths (500 micron diamond)
 
*: radiator thickness: 0.0025 radiation lengths (500 micron diamond)
 
*: max power in photon beam: 4.5 W
 
*: max power in photon beam: 4.5 W
 +
Now one is fully within the limits of the beamline design for GlueX, in terms of photon beam power being incident at the collimator cave.

Revision as of 13:33, 31 October 2012

Running conditions of 2-photon experiment in Hall B

  • based on numbers in this talk by Robert Bennett
    primary beam energy: 5.5 GeV
    max beam current: 120 nA
    radiator thickness: 0.009 radiation lengths
    max power in photon beam: 5.9 W

For comparison, here are the maximum design numbers for polarized running with GlueX. Note that these rates are a bit higher than the standard high-luminosity running conditions, but these are the upper limits that we simulated when we designed the collimator cave shielding and beam line instrumentation.

  • primary beam energy: 12 GeV
    max beam current: 3 μA
    radiator thickness: 0.0001 radiation lengths
    max power in photon beam: 4.8 W

Imagine just taking the running conditions from the TPE experiment in Hall B and replicate them in Hall D, sticking as close as possible to the GlueX photon beam configuration.

  • primary beam energy: 12 GeV
    max beam current: 3 μA
    radiator thickness: 0.00025 radiation lengths (50 micron diamond)
    max power in photon beam: 12 W

The radiator thickness was increased just enough to get the same photon beam intensity as the Hall B experiment. Notice that the total photon beam is only a factor 2.5 higher absorbed dose in the collimator cave than expected for high-luminosity running in GlueX. A more conservative configuration with the same photon beam power might be the following, with the same photon beam intensity.

  • primary beam energy: 12 GeV
    max beam current: 300 nA
    radiator thickness: 0.0025 radiation lengths (500 micron diamond)
    max power in photon beam: 12 W

Another idea to think about:

  • O(m/E) precollimation of photon beam at exit from tagger hall would give factor 2.7 reduction in beam power, also harden the spectrum.
    1m/E collimator @ 30m (radiator - tagger hall exit) has diameter 2.6mm which is feasible in terms of alignment and beam focal spot size. Under these conditions, the following would apply.
    primary beam energy: 12 GeV
    max beam current: 300 nA
    radiator thickness: 0.0025 radiation lengths (500 micron diamond)
    max power in photon beam: 4.5 W

Now one is fully within the limits of the beamline design for GlueX, in terms of photon beam power being incident at the collimator cave.