Fe55 with premixed 50/50 Ar/CO2, New preamp and new HVB, ADC half-range, for a range of HV and (later) gas flow rate
From New HVB:

New preamp and HVB - max amplitude histo for 55Fe
New preamp and HVB - energy histo for 55Fe

Mixed 50/50 at 80 sccm total, less HV

max amplitude histo for 55Fe mixed 50/50 at 80 sccm
energy histo for 55Fe mixed 50/50 at 80 sccm

Mixed 50/50 at 40 sccm total - note the change in x-axis scale. The 2050V run was wrong initially, corrected 1340 Oct 27.

max amplitude histo for 55Fe mixed 50/50 at 40 sccm
energy histo for 55Fe mixed 50/50 at 40 sccm

Various gas flow rates at 1850V (left) and 2050V (right)

energy histo for 55Fe mixed 50/50 at 1850V
energy histo for 55Fe mixed 50/50 at 2050V


Events at 80 sccm total, 1850V (left) and 2050V (right)

energy histo for 55Fe mixed 50/50 at 1850V
energy histo for 55Fe mixed 50/50 at 2050V


Events at 40 sccm total, 1850V (left) and 2050V (right)

energy histo for 55Fe mixed 50/50 at 1850V
energy histo for 55Fe mixed 50/50 at 2050V

As above, zoomed in on pedestal... 80 sccm total, 1850V (left) and 2050V (right)

energy histo for 55Fe mixed 50/50 at 1850V
energy histo for 55Fe mixed 50/50 at 2050V

40 sccm total, 1850V (left) and 2050V (right) The little blip at ~ch 250 in the RH plot is an afterpulse, this is not noticeable at 2000V.

energy histo for 55Fe mixed 50/50 at 1850V
energy histo for 55Fe mixed 50/50 at 2050V


Premixed gas at different flow rates, cylinder -> small flowmeter -> CDC prototype. I used the small flowmeter and the bubble count to very approximately match the flow rates to those from the mixer...

80sccm from the mixer ~ off the top of the small flowmeter ~ 1.9 bubbles/s (+/- 0.2) 40sccm from the mixer ~ small flowmeter measures 57 ~ 1.1 bubbles/s estimated c. 20sccm from the mixer ~ small flowmeter measures 27 ~ 0.25 bubbles/s (+/- 0.02) - the prototype used to bubble at this rate months ago
estimated c. 10sccm from the mixer ~ small flowmeter measures 15 ~ 0.04 bubbles/s (+/- 0.02)
The non-linear correspondence between flowmeter and bubble rate implies that the prototype is leaking at or below 40sccm.


1850V, amplitude (left) and energy (right)

amplitude histo for 55Fe premixed 50/50 at 1850V
energy histo for 55Fe premixed 50/50 at 1850V

1950V, amplitude (left) and energy (right)

amplitude histo for 55Fe premixed 50/50 at 1950V
energy histo for 55Fe premixed 50/50 at 1950V

2050V, amplitude (left) and energy (right)

amplitude histo for 55Fe premixed 50/50 at 2050V
energy histo for 55Fe premixed 50/50 at 2050V