Line 8: Line 8:
  
 
Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below).  Blue histograms are tracked hits, runtimes were not equal.
 
Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below).  Blue histograms are tracked hits, runtimes were not equal.
The fADC range is 0-4096.  These are histograms of the first maximum in ADC value for each event, pedestal has been subtracted.  
+
The fADC range is 0-4096.  These are histograms of the first maximum in ADC value for each event, pedestal has been subtracted, maximum must be above threshold.  Threshold is pedestal mean + 5sigma.  
 
Pedestal mean was approx 100, sigma is 9-12.  
 
Pedestal mean was approx 100, sigma is 9-12.  
  
Line 22: Line 22:
 
[[Image:run_31646_thr5_amp6_unz.png|thumb|420px|Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V]]
 
[[Image:run_31646_thr5_amp6_unz.png|thumb|420px|Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V]]
 
|}
 
|}
 +
 +
Below: threshold pedestal mean + 4sigma
 +
{| border="0" cellpadding="2"
 +
|width="420pt"|
 +
[[Image:run_31646_thr4_amp6.png|thumb|420px|Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V]]
  
 
Earlier new fADC results are on this page: [[New_fADC_with_cosmics]]
 
Earlier new fADC results are on this page: [[New_fADC_with_cosmics]]

Revision as of 10:02, 20 April 2011

  • 50/50 mixed Ar, CO2
  • Recalibrated MFCs
  • Outer plenum added to CDC prototype to prevent leaks
  • New modified HVB
  • 2050V
  • Comparison of modified preamp (preamp4 with terminating resistors removed) and its predecessor (preamp4)


Below: cosmics with new fADC, preamp4 (top) and newest modified preamp (below). Blue histograms are tracked hits, runtimes were not equal. The fADC range is 0-4096. These are histograms of the first maximum in ADC value for each event, pedestal has been subtracted, maximum must be above threshold. Threshold is pedestal mean + 5sigma. Pedestal mean was approx 100, sigma is 9-12.

Max amplitude histo for cosmics, new fADC, modified HVB, mixed 50/50 at 60 sccm 2050V
Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V
Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V

Below: threshold pedestal mean + 4sigma

Max amplitude histo for cosmics, new fADC, modified HVB, modified preamp, mixed 50/50 at 60 sccm 2050V

Earlier new fADC results are on this page: New_fADC_with_cosmics



Below: 55Fe spectra with preamp 4 and Struck ADCs set to half-range, 550mV

Max amplitude histo, 55Fe, 2050V, Struck fADC range 550mV
Max amplitude histo (x-axis expanded), 55Fe, 2050V, Struck fADC range 550mV
Integrated, 96ns, 55Fe, 2050V, Struck fADC range 550mV

Below: 55Fe spectra with newest preamp (preamp4 minus terminating resistors) and Struck ADCs set to full-range, 1.1V

Max amplitude histo, 55Fe, 2050V, Struck fADC range 1.1V
Max amplitude histo (x-axis expanded), 55Fe, 2050V, Struck fADC range 1.1V
Integrated, 96ns, 55Fe, 2050V, Struck fADC range 1.1V