Revision as of 12:45, 28 January 2020
Description on the Symptoms
After restarting DAQ, the missing tiles can be seen on the occupancy plots on Rootspy (see an example image to the right).
Solution
- Stop the current run.
- Completely restart Coda (by pressing STOP button on the RCM GUI).
- Start Coda normally (Configure->Download->Prestart->Start)
- Call DIRC expert if problem was not resolved.
Occupancy of the Issue
Date | " " Resolved After DAQ Reboot " " | Box Tile | Log Post |
---|---|---|---|
2020 Jan 27, 15:18 | Yes | North c1 r15 | https://logbooks.jlab.org/entry/3772139 |