Difference between revisions of "Silicon Photomultiplier Performance"

From GlueXWiki
Jump to: navigation, search
(PDE measurements of Photonique SiPM devices ''(R. Jones)'')
(PDE measurements of Photonique SiPM devices ''(R. Jones)'')
Line 72: Line 72:
 
|}
 
|}
  
The above PED values are for T=+22° and 1.5V over the avalanche threshold.  Measurements performed on the bench at Connecticut have confirmed the manufacturer's PED values at 570 nm, the wavelength of the yellow LED we used for these tests.
+
The above PDE values are for T=+22° and 1.5V over the avalanche threshold.  Measurements performed on the bench at Connecticut have confirmed the manufacturer's PDE values at 570 nm, the wavelength of the yellow LED we used for these tests.
  
Comparing 24% for the 2x2&nbsp;mm<sup>2</sup> Photonique device with 5%
+
If one wants to compare the Sensl and Photonique devices, it is only fair to compare them under the same operating conditions: 1.5V over threshold.  Under these conditions, Carl measured the actual efficiency of the Sensl 1&nbsp;Comparing 24% for the 2x2&nbsp;mm<sup>2</sup> Photonique device with 5%

Revision as of 15:18, 18 April 2007

PDE measurements of Sensl 1mm devices (E. Smith)

The following numbers are taken from the first two rows of Zisis talk at the collaboration meeting (p 5, 1x1 mm2 SiPM parameters)

Design Pixels DR @ 4V Fill Factor Dark Current Peak PDE @ 4V
(MHz) (%) (μA) (%)
C20 620 2.9 17 -2 8.5
A20L 920 5.9 34 -3 17.0

From Carl's talk (p 37, QE and PDE), you can determine the ratio of the PDE@4V to PDE@1.2V, which was the chosen operating point for Carl's measurements. The ratio is approximately 2. Therefore, the expected PDE@1.2V for the C20 is about 8.5/2 = 4.2. Carl measured it to be 4.4 ± 0.5.

Conclusion: At the operating point of about 1.2 V above breakdown, the PDE for the C20 (1mm2, 17% fill factor) SiPM is about 4.5%. So far, our experience tells us that the practical operating point for these devices is about 1V over voltage. As the voltage above the breakdown increases further signal to noise decreases and even pulse shape become distorted.

If we use the A20L instead of the C20, the PDE will double, but so will the dark rate. We believe that the estimated 110  MHz rate for the 4x4 array is for the 1V over voltage, thereby consistent with a PDE of 4.5%.

Comments are welcome. In particular, it would be useful for Richard to remind us the conditions that were used for his measurements of the PDE for the 1x1 and 2x2 mm2 SiPMs.

PDE measurements of Photonique SiPM devices (R. Jones)

The following numbers are taken from the report gluex-doc-760. They are consistent with the data sheets for these devices available on the photonique web site. The SSPM_050701GR_TO18 SiPM (first row in the table) is the 1 mm2 device. It has a slightly higher peak PDE at 600 nm than the newer 4.4 mm2 device (second row in the table) but, but the larger SiPM efficiency drops off more slowly at shorter wavelengths and so it is actually more efficient in the green-blue region.

Device Pixels DR @ 1.5V Peak PDE PDE(500nm)
(MHz) @ 1.5V (%) @ 1.5V (%)
SSPM_050701GR_TO18 556 9.8 30 18
SSPM_0606BG4MM_PCB 1700 8.9 27 24

The above PDE values are for T=+22° and 1.5V over the avalanche threshold. Measurements performed on the bench at Connecticut have confirmed the manufacturer's PDE values at 570 nm, the wavelength of the yellow LED we used for these tests.

If one wants to compare the Sensl and Photonique devices, it is only fair to compare them under the same operating conditions: 1.5V over threshold. Under these conditions, Carl measured the actual efficiency of the Sensl 1 Comparing 24% for the 2x2 mm2 Photonique device with 5%