Difference between revisions of "Spring 2017 High Intensity Run Plan"
From GlueXWiki
(→Analysis Plan) |
(→Run Conditions) |
||
Line 3: | Line 3: | ||
== Run Conditions == | == Run Conditions == | ||
− | * Radiator: JD70-100 (58 μm - 4.8·10<sup>-4</sup> R.L., 7×7 mm²) | + | * Radiator: JD70-100 (58 μm - 4.8·10<sup>-4</sup> R.L., 7×7 mm²) and/or 3x10^-4 Al radiator |
* CW beam current up to 500 nA | * CW beam current up to 500 nA | ||
* Tagger quadrupole on (-4.2 A). | * Tagger quadrupole on (-4.2 A). | ||
* 5.0 mm collimator hole | * 5.0 mm collimator hole | ||
* Detector Settings | * Detector Settings | ||
− | ** CDC | + | ** CDC: 2125V HV, with following trip settings: |
+ | *** 24 uA for 300 nA beam | ||
+ | *** 36 uA for 500 nA beam | ||
** FDC and TOF are fine with nominal settings | ** FDC and TOF are fine with nominal settings | ||
** Lower TAGH counters remain off | ** Lower TAGH counters remain off |
Revision as of 18:54, 6 March 2017
This is the plan for running high intensity tests for the Spring 2017 run
Run Conditions
- Radiator: JD70-100 (58 μm - 4.8·10-4 R.L., 7×7 mm²) and/or 3x10^-4 Al radiator
- CW beam current up to 500 nA
- Tagger quadrupole on (-4.2 A).
- 5.0 mm collimator hole
- Detector Settings
- CDC: 2125V HV, with following trip settings:
- 24 uA for 300 nA beam
- 36 uA for 500 nA beam
- FDC and TOF are fine with nominal settings
- Lower TAGH counters remain off
- CDC: 2125V HV, with following trip settings:
- Trigger settings
- Standard triggers
- Include BCAL+FCAL+ST?
- Prescale factor: 3?
- Standard triggers
Run Plan
- 2 current points: 300 nA and 500 nA
- present data at 150 nA used as reference
- 2? runs: 1 PARA, 1 PERP
Analysis Plan
- Study rho, omega rate
- Statistical precision of asymmetry
- 1 run will give rho asymmetry on level of 40 +- 1 %
- omega statistics are factor of ~10 less
- Estimate fraction of accidentals?