After PMT leads and cables have been soldered to PCBs:  
 
After PMT leads and cables have been soldered to PCBs:  

Revision as of 21:30, 29 September 2013

PSL in light tight box.png

After PMT leads and cables have been soldered to PCBs:

  • Place assembled PMTs in light-tight box and connect cables
  • Verify that PMT, diffuser & attenuator stand, and Picosecond Laser are aligned
  • Turn on HV and check monitored current, ramp up to 850 V
  • Turn on Picosecond Laser (intensity scale to 1.5, freq. to 100 kHz)
  • Check signals on oscilloscope, use measurement function to report min/max within the peak and integral
  • Ramp laser frequency to 1 MHz, 2 MHz, 5 MHz, 10 MHz
  • Repeat frequency scan for HV=900 V, 1000 V, 1100 V
  • Check for break-down voltage and frequency
  • Report results on excel sheet, save snapshots of scope signal