Difference between revisions of "Forward Drift Chamber Shift"

From Hall D Ops Wiki
Jump to: navigation, search
(Routine operation)
Line 23: Line 23:
 
The most important task for the shift workers is to monitor the parameters (using GUIs) of the following FDC related subsystems: High Voltage, Low Voltage (including thresholds), Gas system, and the Cooling system.  A description of how to bring up the corresponding GUI screens and how to operate them can be found in [[Slow Controls Shift| Section Slow Controls]]. The HV and LV subsystems, as well as and the Gas and Cooling systems have common GUIs. The GUIs are designed to show green color if all the parameters in a certain subsystem/subgroup are within the limits. If any of the parameters is out of limits the background for the corresponding subsystem/subgroup will turn red. Examples of the GUIs are shown below:
 
The most important task for the shift workers is to monitor the parameters (using GUIs) of the following FDC related subsystems: High Voltage, Low Voltage (including thresholds), Gas system, and the Cooling system.  A description of how to bring up the corresponding GUI screens and how to operate them can be found in [[Slow Controls Shift| Section Slow Controls]]. The HV and LV subsystems, as well as and the Gas and Cooling systems have common GUIs. The GUIs are designed to show green color if all the parameters in a certain subsystem/subgroup are within the limits. If any of the parameters is out of limits the background for the corresponding subsystem/subgroup will turn red. Examples of the GUIs are shown below:
  
[[Image:BCAL_environmental_sensors.png | thumb | 400px | Fig. 1. Environmental sensors for the BCAL. The red ovals highlight the information provided to the user, available for the upstream and downstream ends independently.
+
[[Image:FDC_GAS_system_screen.jpg | thumb | 400px | Fig. 1. ]]
This includes the forty-eight temperature readings of each RTD on the cooling plates, which are in thermal contact with the back of the MPPC sensors, the nitrogen flow, and the four
+
measurements of temperature, humidity and dew point  of the gaseous atmosphere of the readout electronics.]]
+
  
 
    
 
    

Revision as of 16:52, 2 June 2014

The Forward Drift Chambers

The Forward Drift Chamber (FDC) system consists of four identical packages, each having six chambers (cells). These are planar drift chambers with both wire and cathode strips being readout, in total 12,672 channels.

The cells within a package have independent gas volumes, but are separated with a flexible mylar membrane and therefore the pressure difference between neighboring cells should never exceed 30~Pa and should normally be less than 5~Pa.

Positive (up to 2200V) and negative (up to 500V) HV is applied on the sense and field wires respectively with dark currents not exceeding 3 microA per HV channel .

The detector (including cables) emits a total power of about 1500~Watt, of which about 900~Watt inside the magnet, due to the LV applied on the detector pre-amplifieres. A cooling system using Fluorinert is used to keep the temperature on the pre-amplifiers within the permissible limits. The LV system should never be turned ON if the cooling system doesn't work .

Routine operation

The most important task for the shift workers is to monitor the parameters (using GUIs) of the following FDC related subsystems: High Voltage, Low Voltage (including thresholds), Gas system, and the Cooling system. A description of how to bring up the corresponding GUI screens and how to operate them can be found in Section Slow Controls. The HV and LV subsystems, as well as and the Gas and Cooling systems have common GUIs. The GUIs are designed to show green color if all the parameters in a certain subsystem/subgroup are within the limits. If any of the parameters is out of limits the background for the corresponding subsystem/subgroup will turn red. Examples of the GUIs are shown below:

Fig. 1.