Difference between revisions of "Solenoid Ramping Shift"

From Hall D Ops Wiki
Jump to: navigation, search
(Created page with "== The Solenoid Magnet == The principle analysis magnet for the GlueX spectrometer is a superconducting solenoid made from the Large Aperture Solenoid Spectrometer (LASS) sol...")
 
(Setup)
(45 intermediate revisions by 3 users not shown)
Line 1: Line 1:
== The Solenoid Magnet ==
 
  
The principle analysis magnet for the GlueX spectrometer is a superconducting solenoid
 
made from the Large Aperture Solenoid Spectrometer (LASS) solenoid built at
 
Stanford Linear Accelerator Center (SLAC) in 1971.
 
The LASS solenoid consists of four separate coils located in separated vacuum vessels.
 
Three coils were later used a second time for the Muon decays into Electron and GAmma ray (MEGA) experiment
 
at Los Alamos National Lab from 1993 through 1995.
 
During the MEGA experiment, the coils had electrical shorts and leaks to the insulating vacuum
 
along with deteiorated superinsulation and instrumentation.
 
In order to use them for the GlueX experiment, all four LASS coils have been extensively repaired,
 
refurbished and tested at Jefferson Lab.
 
The four coils were then assembled in Hall-D.
 
The final GlueX solenoid has a bore of 1.85~m and measures 4 m long.
 
 
[[File:halld_solenoid.jpg | 400px ]]
 
 
The magnet consists of 4608 windings(turns) of conductor in total.
 
The windings are grouped into 19 sub-coils and distributed in four coils.
 
The conductor is made of two copper bars 7.62 mm wide and about 2 mm thick,
 
with a band of about 1~mm thick NbTi superconductor composite soldered in between them.
 
These double-pancake windings are not insulated at their sides and are separated by a 0.64 mm thick stainless steel band
 
co-wound with the cable, and insulated from the cable on both sides by Mylar bands of 0.2 mm thick.
 
 
[[File:halld_solenoid_simulation.png | 600px ]][[File:halld_solenoid_field.png | 500px ]]
 
  
 +
== Routine Operation ==
 
When powered with the maximum current of 1350 A, the solenoid generates a field close to 2 Tesla in the bore.
 
When powered with the maximum current of 1350 A, the solenoid generates a field close to 2 Tesla in the bore.
 
The full inductance of the solenoid is about 26 H.
 
The full inductance of the solenoid is about 26 H.
 
+
As the superconducting solenoid is very complex system, the operation of such system requires considerable knowledge and experience. The commissioning, troubleshooting, tuning of ground fault and quench detectors, Power ON and OFF, change current will be conducted by authorized experts only.
== Routine Operation ==
+
 
+
As the superconducting solenoid is very complex system, the operation of such system requires a lot of knowledge and experiences. The commissioning, troubleshooting, tuning of ground fault and quench detectors, Power ON and OFF, change current will be conducted by authorized experts only.
+
  
 
The responsibilities of the shift workers include:
 
The responsibilities of the shift workers include:
Line 39: Line 13:
 
* Make log entries to the [https://logbooks.jlab.org/book/HDLOG HDLOG] and [https://logbooks.jlab.org/book/HDSOLENOID HDSOLENOID] logbooks.
 
* Make log entries to the [https://logbooks.jlab.org/book/HDLOG HDLOG] and [https://logbooks.jlab.org/book/HDSOLENOID HDSOLENOID] logbooks.
  
== How-To ==
+
These procedures are documented in the [https://misportal.jlab.org/jlabDocs/document.seam?id=85132 Hall D Solenoid Normal Ramp up Procedure for Shift takers (D00000-04-02-P013)] . Additional information can be found in detailed [[Solenoid Operations | Operation Instructions]]  and  [[Solenoid_Expert | Expert Information]].
  
=== Open Solenoid EPICS Control GUIs ===
+
== Ramping the Magnet ==
 +
[[Image:Solenoid Vacuum Status.png | thumb | 485px | Fig. 1. Solenoid vacuum status screen]]
 +
[[Image:Solenoid status pickup coil.png | thumb | 485px | Fig. 2. Solenoid pickup coil status screen]]
 +
[[Image:Solenoid status forces.png | thumb | 485px | Fig. 3. Solenoid strain gauges]]
  
The Solenoid Main GUI can be opened from the Hall-D's main EPICS GUI.
+
'''This procedure is to be used by shift operators after all the preliminary checks have been completed and approval to ramp has been given by the Solenoid system expert.''' It assumes the magnet is at 0A current. The slew rates to ramp the magnet are defined in the state machine and are
 +
0.2A/s up to 800A, 0.12A/s from 800 to 1000A and 0.06A/s after that.
  
If not opened, follow these steps:
 
  
# If in counting house, go to step 3 directly. If offsite or using a person computer, log on to JLab's Login Gateway: <code>ssh -X yourusername@login.jlab.org</code>
+
=== Setup ===
# If not already in the device network, log on to the Hall Gateway (hallgw): <code>ssh -X yourusername@hallgw</code>, use your two-step crytocard to generate passcodes.
+
# Pull up CSS Solenoid screens as follows:
# Log on to one of the gluon machine '''gluon01''' - '''gluon05''', '''gluon27'''-'''gluon30''', with account hdops: <code>ssh -X hdops@gluon''X''</code>, ask Yi Qiang [mailto:yqiang@jlab.org] or Hovanes Egiyan [mailto:hovanes@jlab.org] for password.
+
#* Pull up Solenoid Control screen from CSS main screen - Magnet/Solenoid
# Type <code>gluex_css</code> to bring up the CSS framework. Browse and open one of the solenoid profiles: '''Solenoid''' or '''Solenoid''n''''', where ''n'' = 2 - 5.
+
#* Cryo System (from Solenoid Control screen)
# Switch to '''OPI Runtime''' perspective if not already, [[Image:Solenoid_Screen_Switch_OPI.png | 200px]].
+
#* Interlock
# Open '''Main Action Bar''' using short cut [[Image:Solenoid_Screen_shortcut.png | 200px]]
+
#* Power Supply
# From '''Main Action Bar''', click the '''Solenoid''' button in section '''MAGNETS''' to bring up the Solenoid Main GUI.
+
#* Voltage Taps/PXI
 +
#* Vacuum Pumps
 +
#* Strain Gauges
 +
# From the “Voltage Taps/PXI fast DAQ" screen, do the following;
 +
#* Click on the “Plot all Pickup Coils” button to open a Voltage vs. Time graph
 +
#*      Stop Recording Root File and Stop the PXI DAQ
 +
#*      Press the Calibrate button and wait until voltages zero out
 +
#* Ensure the PXI DAQ is running (if not click “Start PXI DAQ” button
 +
#* Click on “Start Recording ROOT File” button. Make sure “File Size” is increasing.
 +
#* Click on the “Current” box at the top to start a Current vs. Time graph
 +
# On the “Solenoid Strain Gauges” window:
 +
#* Confirm magnet is at zero current. Click on the “Zero Strain Gauges” button (all strain gauges should read close to 0).  
 +
#* Click on “Plot All Forces” button (Graph for 4 coils should be displayed)
 +
#* Close strain gauge window
 +
# Ensure all is green on the “Vacuum Status” window and pumps are at 100% or better. The screen should look like in Fig. 1.
 +
# On “Interlock Status” screen, all should be green except the Main Contactor which will be red until the MPS is turned on. If not, click on the “Reset Interlocks” button.  
 +
# Confirm Reset worked. If not, try again. If it still did not work, contact solenoid expert.
 +
# Arrange all the screens so they are visible.
  
Basic status of the power supply will be shown on the main menu,  in order to access detailed information and functions, click the corresponding buttons.
+
===Ramp up===
  
Screenshots of all GUI pages can be found in [[Solenoid_EPICS_Screen]].
 
  
''note'': [http://mobaxterm.mobatek.net/download-home-edition.html MobaXterm] is recommended for Windows users, and a copy is available on M: drive <code>M:\halld-online\MobaXterm.exe</code>
+
# On “MPS Auto Control” screen, all should be green except “MPS On” box. If not, click on the “Reset MPS” button.
 +
# Confirm Reset (all boxes except “MPS On” are green).
 +
# Set current in “Final Goal” box to 0 (zero) and hit enter.
 +
# Confirm 0 (zero) reading in “Final Goal” box.
 +
# Click on the “Turn On MPS” button. Confirm all boxes are green.
 +
# Confirm MPS is on (all boxes green).
 +
# Set desired current in the “Final Goal” box.
 +
# Click on the ramp button. Note: the ramps are programmed into the system and will soak at 800A for 15 minutes, 1000A for 1 hr and 1200A for 1 hr
 +
# Confirm magnet is ramping by seeing an increase in current.
 +
# Pay close attention to the “Pick Up Coil” and “Strain Gauge” graphs to ensure the signals are acting normally (see Figs. 2 and 3). If not, click on the “Cancel” button and notify solenoid expert.
 +
# When the current is at approximately 40A, take a screen shot of all the windows and post an HDLOG and HDSOLENOID log.
 +
#* Press Alt+Prt Scrn
 +
#* Either pick “current window” or “all windows” (if you pick current window, only the active window will be used and you must repeat for all windows) and click ok
 +
#* Add “solenoid ramp” (or specific window) to the name of the file after the date and save. The correct folder is hdops/pictures/solenoid screenshots and pictures.
 +
#* Make an HDLOG/HDSOLENOID entry and attach screenshots
 +
# If there is a dump, ensure an HDLOG is made with a current screen shot and contact mechanical on call. Do not reset anything until instructed to do so.
 +
# If a ramp is cancelled/stopped for any reason and the system expert gives the ok to continue ramping, type in the new “Final goal” set point and click the ramp button.
  
[[Image:Solenoid_Screen_Main_Action_Bar.png|x300px]]
+
===Ramp down===
[[Image:Solenoid_Screen_Main.png|x300px]]
+
# To ramp down to zero current, type in “0” as the “Final Goal” and click the ramp button.
 +
# Take screenshots of any relevant windows, especially if there are any anomalies during rampdown.
  
=== Open EPICS Alarm GUI ===
+
== When the Solenoid Trips ==
  
The EPICS Alarm GUI is based on the same framework of the control GUIs. One can easily switch to the '''Alarm''' perspective to bring up the Alarm Handler of the whole Hall-D system.
+
* Call Mechanical On-Call (757-544-7258) every time when the power supply trips!
[[Image:Solenoid_Screen_Switch_Alarm.png | 200px]]
+
* DO NOT clear interlock status or attempt to change MPS state before the Solenoid On-Call responses!
 +
* Make screen shots of Solenoid status and post them to the [https://logbooks.jlab.org/book/HDLOG HDLOG] and [https://logbooks.jlab.org/book/HDSOLENOID HDSOLENOID] logbooks: Interlock, MPS, Voltage, Cryo, Vacuum, Strain Gauge.
  
[[Image:Solenoid_Screen_Alarm.png|400px]]
 
  
The list of alarm thresholds is presented in [[Media:Solenoid_Alarm.pdf]].
+
== PXI DATA Monitoring ==
 +
There are two applications that allow you to view the PXI data "analyzer" and "pximon".  
  
=== Ramp the Magnet ===
+
=== PXI Data Analyzer ===
 +
The main diagnostics tool to view all PXI voltage tabs is called "analyzer". It can be started directly from the PXI volage tabs GUI clicking on the button "Analyzer" or by going to the directory "/home/hdops/solenoid/" and type in there "./analyzer". Using this GUI you can open any PXI data file that contains all data of the voltage tabbs, pick-up coils and accelerometers at a 10kHz sample rate (see fig. 1).
 +
[[Image:AnalyzerPXIdata.png | thumb | 485px | Fig. 1. PXI Data Analyzer]]
  
Special care is need to prevent sudden change of the solenoid current from damaging the magnet. And the solenoid current is ramped with stops and various ramp rates:
+
=== Accelerometer Monitor ===
 +
To monitor the accelerometeres start the program in "/home/hdops/solenoid" with the command "./pximon_1". This will start up a GUI in which you will click the button "Startthreads". This will show the accelerometer data for each coil separately that the user can select with the raido buttons at the bottom of the GUI. (see fig. 2)
 +
[[Image:accelerometerMonitor.png | thumb | 485px | Fig. 2. Accelerometer Monitor]]
  
* 0 - 800 A: 0.2 A/s
 
* sit at 800 A for 15 minutes
 
* 800 - 1000 A: 0.12 A/s
 
* sit at 1000 A for 60 minutes
 
* 1000 - 1200 A: 0.06 A/s
 
* sit at 1200 A for 60 minutes
 
* 1200 A - beyond: 0.06 A/s
 
  
The ramp down follows the reversed procedure.
 
 
Such a stepped ramp profile is automatically executed by an EPICS State Machine through EPICS GUI. To ramp the magnet, follow the steps:
 
 
# Check the status of the Solenoid including cryo and vacuum, make sure all interlocks except for the MPS main contactor are OK.
 
# Open the MPS GUI by clicking '''Power Supply''' button in the main menu.
 
# Verify the MPS is in Remote mode. If not, click the '''Set Remote''' button.
 
# If the MPS is not ON, turn it ON by clicking the '''Turn ON MPS''' button.
 
# Type the desired current to field next to '''Final Goal'''.
 
# Click '''Ramp''' button to start a ramp. The button should then switch to '''Cancel''' and the status indicator will tell you that a ramp is in progress.
 
# Click '''Cancel''' button to abort a ramp.
 
 
For rare cases, one may want to ramp the magnet using manual settings
 
 
# Open MPS Manual Control by clicking the '''Manual''' button.
 
# Type in desired current and ramp rate in the fields next to '''Set Point'''.
 
# Click '''Ramp''' to start a ramp.
 
# Click '''Cancel''' to abort a ramp.
 
# Click '''Auto''' button to go back to Automatic Mode.
 
 
=== Take High Speed Data using PXI ===
 
 
High speed data are needed to trouble shoot issues associated with the solenoid. The PXI ROOT file recorder needs to be turn on every time the magnet is being ramped. To do so
 
 
# Open the '''Voltage Taps/PXI''' GUI from the main menu.
 
# Verify that the PXI status shows '''DAQ Running'''.
 
# Click '''Start Recording ROOT File''' button to start a writing session. One will notice a new file is generated and the file size is increasing.
 
# Click '''Stop Recording ROOT File''' button to stop a writing session after the solenoid is settled.
 
# One can check the recorded data using the '''Analyzer'''.
 
 
More information of PXI can be found in [[Solenoid_expert_pxi | expert page]].
 
 
=== Reboot EPICS IOCs ===
 
 
There are two IOCs associated with the Solenoid operation:
 
 
; Solenoid PXI Interface: IOC to record high speed data from the PXI
 
; Solenoid PLC Interface: IOC to pass tags between PLC and EPICS
 
 
One can reboot eith IOC following the procedure listed [[Slow_Controls_Shift#Input.2FOutput_Controllers_.28IOC-s.29 | here]]
 
 
Every time after Solenoid PLC Interface is rebooted, one needs to reload the alarm setting from the '''Interlock''' GUI to store all the alarms.
 
 
# Open the '''Interlock''' GUI from the Main Menu.
 
# Click '''Restore Alarm Settings'''.
 
# Select <code>alarm_commissioning.snap</code> from the file list and click '''Restore'''.
 
 
=== Access PLC HMI Interface ===
 
 
The HMI screens are for expert only. '''halldsc7''' currently serves as the Remote Desktop Server to access these screens. Right now, only ONE connection is allow at a time, so please quit the remote desktop if not needed. You can access them by using any Linux terminals in the Counting House: gluon01a, gluon02, gluon03, gluon04 and gluon5.
 
To do so, follow theses steps:
 
# Open a terminal from any of the '''gluon''' machines.
 
# Start remote desktop in fullscreen mode by typing <code>rdesktop -f halldsc7</code> in the terminal
 
# In the Windows Log In screen, choose '''Other User''', use '''JLAB\halldusers''' for username, ask Yi [mailto:yqiang@jlab.org] or Hovanes [mailto:hovanes@jlab.org] if you do not know the password.
 
# Once logged in, double click the '''FactoryTalk Client''' shortcut on the desktop, choose '''...\Hall_D.cli''' application and click '''run''' to start the HMI screens.
 
# Once done, please log off the remote desktop. You can leave the screens running though.
 
 
More information about HMI screens can be found in [[Solenoid_expert_plc | expert page]].
 
 
=== When a Trip Happens ===
 
 
* Call Mechanical On-Call (757-544-7258) every time when the power supply trips!
 
* DO NOT clear interlock status or attempt to change MPS state before the Solenoid On-Call responses!
 
* Make screen shots of Solenoid status and post them to the [https://logbooks.jlab.org/book/HDLOG HDLOG] and [https://logbooks.jlab.org/book/HDSOLENOID HDSOLENOID] logbooks: Interlock, MPS, Voltage, Cryo, Vacuum, Strain Gauge.
 
 
Only when instructed by the expert, follow the following steps to turn MPS back ON:
 
 
# Clear interlock by clicking '''Reset Interlocks''' button from the Interlock screen.
 
# Verify that the MPS is in <code>REMOTE</code> mode from the MPS screen. If not, click '''Set Remote''' button to switch.
 
# Clear MPS interlock by clicking '''Reset MPS''' button. Second attempt may be needed to clear all the error bits.
 
# Turn on the power supply by clicking '''Turn ON MPS''' button.
 
  
 
== Expert personnel ==
 
== Expert personnel ==
Line 167: Line 104:
 
! width=200px | Name  !!  width=100px | Extension !!  width=100px | Cellphone !!  width=100px | Pager !! Date of qualification  
 
! width=200px | Name  !!  width=100px | Extension !!  width=100px | Cellphone !!  width=100px | Pager !! Date of qualification  
 
|-
 
|-
| Mechanical On-Call ||  align=center |  ||  align=center | 544-7258 ||  align=center |  ||  align=center | Nov 10, 2014
+
| Mechanical On-Call ||  align=center |  ||  align=center | 544-7258 ||  align=center |  ||  align=center | Sept 19, 2017
 
|-
 
|-
| George Biallas - Solenoid ||  align=center | 269-7535 ||  align=center | 813-1179 ||  align=center | [mailto:biallas-page@jlab.org] ||  align=center | June 12, 2014
+
| Tim Whitlatch- Solenoid ||  align=center | 269-5087 ||  align=center | 876-1766 ||  align=center | [mailto:whitey-page@jlab.org] ||  align=center | Sept 19, 2017
 
|-
 
|-
| Yi Qiang - Solenoid/PXI  ||  align=center | 269-7237 ||  align=center | 358-2146 ||  align=center | [mailto:yqiang-page@jlab.org] ||  align=center | June 12, 2014
+
| Beni Zihlmann - Solenoid/PXI  ||  align=center | 269-5310 ||  align=center | 926-9671 ||  align=center | [mailto:zihlmann-page@jlab.org] ||  align=center | Sept 19, 2017
 
|-
 
|-
| Mark Stevens - Power Supply ||  align=center | 269-6383 ||  align=center | 318-1576 ||  align=center | [mailto:stevensm-page@jlab.org] ||  align=center | June 12, 2014
+
| Mark Stevens - Power Supply ||  align=center | 269-6383 ||  align=center | 310-1576 ||  align=center | [mailto:stevensm-page@jlab.org] ||  align=center | Sept 19, 2017
 
|-
 
|-
| Scot Spiegel - MPS/Solenoid ||  align=center | ||  align=center | 536-5900 ||  align=center | [mailto:spiegel-page@jlab.org] ||  align=center | June 12, 2014
+
| Scot Spiegel - MPS/Solenoid ||  align=center | 269-5900||  align=center | 536-5900 ||  align=center | [mailto:spiegel-page@jlab.org] ||  align=center | Sept 19, 2017
 
|-
 
|-
| Jonathan Creel - Cryogenics ||  align=center | 269-5925 ||  align=center | 869-8910 ||  align=center | [mailto:creel-page@jlab.org] ||  align=center | June 12, 2014
+
| Chris Perry - Cryogenics ||  align=center | 269-6157||  align=center | 371-4926 ||  align=center | [mailto:creel-page@jlab.org] ||  align=center | Sept 19, 2017
 
|-
 
|-
| Dave Butler - PLC  ||  align=center | 329-2583 || align=center | 329-2583 || align=center | [mailto:butler-page@jlab.org] || align=center | June 12, 2014
+
| Jonathan Creel - Cryogenics ||  align=center | 269-5925|| align=center | 869-8910 || align=center | [mailto:creel-page@jlab.org] || align=center | Sept 19, 2017
 
|-
 
|-
| Carroll Jones - LCW  ||  align=center | 876-1778 || align=center | 876-1778 || align=center | [mailto:jonesc-page@jlab.org] || align=center | Nov 14, 2014
 
|}
 
  
== [[Solenoid_Expert | Expert Information]] ==
+
| Nick Sandoval - PLC  ||  align=center | 269-6506 || align=center | (540) 907-2637 || align=center | [mailto:butler-page@jlab.org] || align=center | Sept 19, 2017
 +
|-
 +
| Carroll Jones - LCW  ||  align=center | 876-1778 || align=center | 876-1778 || align=center | [mailto:jonesc-page@jlab.org] || align=center | Sept 19, 2017
 +
|}

Revision as of 08:49, 9 December 2019


Routine Operation

When powered with the maximum current of 1350 A, the solenoid generates a field close to 2 Tesla in the bore. The full inductance of the solenoid is about 26 H. As the superconducting solenoid is very complex system, the operation of such system requires considerable knowledge and experience. The commissioning, troubleshooting, tuning of ground fault and quench detectors, Power ON and OFF, change current will be conducted by authorized experts only.

The responsibilities of the shift workers include:

  • Monitor the overall status of the solenoid: current, field, interlocks, voltages, cryogenic levels etc.
  • Notify experts when anomalies happen.
  • Ramp the magnet per request of run coordinator.
  • Make log entries to the HDLOG and HDSOLENOID logbooks.

These procedures are documented in the Hall D Solenoid Normal Ramp up Procedure for Shift takers (D00000-04-02-P013) . Additional information can be found in detailed Operation Instructions and Expert Information.

Ramping the Magnet

Fig. 1. Solenoid vacuum status screen
Fig. 2. Solenoid pickup coil status screen
Fig. 3. Solenoid strain gauges

This procedure is to be used by shift operators after all the preliminary checks have been completed and approval to ramp has been given by the Solenoid system expert. It assumes the magnet is at 0A current. The slew rates to ramp the magnet are defined in the state machine and are 0.2A/s up to 800A, 0.12A/s from 800 to 1000A and 0.06A/s after that.


Setup

  1. Pull up CSS Solenoid screens as follows:
    • Pull up Solenoid Control screen from CSS main screen - Magnet/Solenoid
    • Cryo System (from Solenoid Control screen)
    • Interlock
    • Power Supply
    • Voltage Taps/PXI
    • Vacuum Pumps
    • Strain Gauges
  2. From the “Voltage Taps/PXI fast DAQ" screen, do the following;
    • Click on the “Plot all Pickup Coils” button to open a Voltage vs. Time graph
    • Stop Recording Root File and Stop the PXI DAQ
    • Press the Calibrate button and wait until voltages zero out
    • Ensure the PXI DAQ is running (if not click “Start PXI DAQ” button
    • Click on “Start Recording ROOT File” button. Make sure “File Size” is increasing.
    • Click on the “Current” box at the top to start a Current vs. Time graph
  3. On the “Solenoid Strain Gauges” window:
    • Confirm magnet is at zero current. Click on the “Zero Strain Gauges” button (all strain gauges should read close to 0).
    • Click on “Plot All Forces” button (Graph for 4 coils should be displayed)
    • Close strain gauge window
  4. Ensure all is green on the “Vacuum Status” window and pumps are at 100% or better. The screen should look like in Fig. 1.
  5. On “Interlock Status” screen, all should be green except the Main Contactor which will be red until the MPS is turned on. If not, click on the “Reset Interlocks” button.
  6. Confirm Reset worked. If not, try again. If it still did not work, contact solenoid expert.
  7. Arrange all the screens so they are visible.

Ramp up

  1. On “MPS Auto Control” screen, all should be green except “MPS On” box. If not, click on the “Reset MPS” button.
  2. Confirm Reset (all boxes except “MPS On” are green).
  3. Set current in “Final Goal” box to 0 (zero) and hit enter.
  4. Confirm 0 (zero) reading in “Final Goal” box.
  5. Click on the “Turn On MPS” button. Confirm all boxes are green.
  6. Confirm MPS is on (all boxes green).
  7. Set desired current in the “Final Goal” box.
  8. Click on the ramp button. Note: the ramps are programmed into the system and will soak at 800A for 15 minutes, 1000A for 1 hr and 1200A for 1 hr
  9. Confirm magnet is ramping by seeing an increase in current.
  10. Pay close attention to the “Pick Up Coil” and “Strain Gauge” graphs to ensure the signals are acting normally (see Figs. 2 and 3). If not, click on the “Cancel” button and notify solenoid expert.
  11. When the current is at approximately 40A, take a screen shot of all the windows and post an HDLOG and HDSOLENOID log.
    • Press Alt+Prt Scrn
    • Either pick “current window” or “all windows” (if you pick current window, only the active window will be used and you must repeat for all windows) and click ok
    • Add “solenoid ramp” (or specific window) to the name of the file after the date and save. The correct folder is hdops/pictures/solenoid screenshots and pictures.
    • Make an HDLOG/HDSOLENOID entry and attach screenshots
  12. If there is a dump, ensure an HDLOG is made with a current screen shot and contact mechanical on call. Do not reset anything until instructed to do so.
  13. If a ramp is cancelled/stopped for any reason and the system expert gives the ok to continue ramping, type in the new “Final goal” set point and click the ramp button.

Ramp down

  1. To ramp down to zero current, type in “0” as the “Final Goal” and click the ramp button.
  2. Take screenshots of any relevant windows, especially if there are any anomalies during rampdown.

When the Solenoid Trips

  • Call Mechanical On-Call (757-544-7258) every time when the power supply trips!
  • DO NOT clear interlock status or attempt to change MPS state before the Solenoid On-Call responses!
  • Make screen shots of Solenoid status and post them to the HDLOG and HDSOLENOID logbooks: Interlock, MPS, Voltage, Cryo, Vacuum, Strain Gauge.


PXI DATA Monitoring

There are two applications that allow you to view the PXI data "analyzer" and "pximon".

PXI Data Analyzer

The main diagnostics tool to view all PXI voltage tabs is called "analyzer". It can be started directly from the PXI volage tabs GUI clicking on the button "Analyzer" or by going to the directory "/home/hdops/solenoid/" and type in there "./analyzer". Using this GUI you can open any PXI data file that contains all data of the voltage tabbs, pick-up coils and accelerometers at a 10kHz sample rate (see fig. 1).

Fig. 1. PXI Data Analyzer

Accelerometer Monitor

To monitor the accelerometeres start the program in "/home/hdops/solenoid" with the command "./pximon_1". This will start up a GUI in which you will click the button "Startthreads". This will show the accelerometer data for each coil separately that the user can select with the raido buttons at the bottom of the GUI. (see fig. 2)

Fig. 2. Accelerometer Monitor


Expert personnel

The individuals responsible for checking that the solenoid is ready to take data and setting its operating parameters are shown in following table. Problems with normal operation of the solenoid should be referred to those individuals and any changes to their settings must be approved by them. Additional experts may be trained by the system owner and their name and date added to this table.

Table: Expert personnel for the Solenoid Magnet
Name Extension Cellphone Pager Date of qualification
Mechanical On-Call 544-7258 Sept 19, 2017
Tim Whitlatch- Solenoid 269-5087 876-1766 [1] Sept 19, 2017
Beni Zihlmann - Solenoid/PXI 269-5310 926-9671 [2] Sept 19, 2017
Mark Stevens - Power Supply 269-6383 310-1576 [3] Sept 19, 2017
Scot Spiegel - MPS/Solenoid 269-5900 536-5900 [4] Sept 19, 2017
Chris Perry - Cryogenics 269-6157 371-4926 [5] Sept 19, 2017
Jonathan Creel - Cryogenics 269-5925 869-8910 [6] Sept 19, 2017
Nick Sandoval - PLC 269-6506 (540) 907-2637 [7] Sept 19, 2017
Carroll Jones - LCW 876-1778 876-1778 [8] Sept 19, 2017