CDC 60 40

From GlueXWiki
Jump to: navigation, search

60/40 Ar/CO2 and cosmics, prototype horizontal

Short runs for range of HV, check pedestal is ok and find MPV

1800V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
1900V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
1950V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma
1975V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma
2000V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma

Fit max amplitude histograms and find optimal HV

60/40 gas, prototype horizontal, 4sigma (gray) and 5sigma (blue) hit threshold, fit range 6sigma+
1950V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (64, 79, 110)
1975V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (63, 84, 124)
2000V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (75, 101, 154)


60/40 gas, noise std dev 16.5, hit threshold 5x16.5 = 82.5, fit range 6pedestalsigma+
HV Fit MPV Fit sigma (FWHM/4) MPV/sigma 99% point 4045/MPV
1950 185 50 3.7 79 22
1975 226 67 3.4 84 18
2000 285 87 3.3 101 14


60/40 gas, prototype horizontal, 4sigma (gray) and 5sigma (blue) hit threshold, fit range 5sigma+
1950V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (56, 72, 103)
1975V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (51, 72, 114)
2000V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (68, 95, 148)


60/40 gas, prototype horizontal, 4sigma (gray) and 5sigma (blue) hit threshold, fit range 5sigma+
1975V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (62, 83, 126)


60/40 gas, noise std dev 16.5, hit threshold 5x16.5 = 82.5, fit range 5pedestalsigma+
HV Fit MPV Fit sigma (FWHM/4) Fit 4sigma/MPV MPV/sigma 4045/MPV 99% point 97% point 95% point
1950 180 51 1.14 3.5 22 72 91 103
1975 219 70 1.28 3.1 18 72 98 114
2000 282 89 1.26 3.2 14 94 128 148
same again, ran for longer
HV Fit MPV Fit sigma (FWHM/4) MPV/sigma 99% point 4045/MPV
1975 233 71 3.3 83 17


Fit drift time histogram

Drift time for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=0
1975V, hit threshold 5sigma, time thresholds 4 and 1 sigma Estimate of t0 is 88 and tmax is 661, difference is 573, straight-line intercept is 81.7, to match histogram fit t0, use tz_add 6.4

Tracking efficiency plot

Tracks/Triggers, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(5sig)-17
tzero=88ns, early hits ignored


Look at position resolution, find best gas mix for Garfield x-t tables

Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=6, 60/40 Garfield (ignore the title saying 50/50, I forgot to change that part of the root script)
1975V, resolution ch 20
1975V, resolution vs hit radius

This does not look nearly as good as 50/50 gas mix. Resolution is calculated as fitted drift distance - measured (interpolated) distance, so it is positive when the measured drift time and distance are too small. In this case they are slightly too small at small drift distance but too large at large drift distance. This could be a problem with the gas mix. For reference, plot res vs distance for tz=5,7,and 0.


Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=8, 60/40 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius

Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times. This would probably have been obvious from the tz=0 data (below) where all the measured drift distances are too large. "Measured" drift distances are actually interpolated from a table of distance calculated for a given gas mix and a range of time measurements. So for this data I should use Garfield tables for a gas mix with slower drift velocity, ie less Ar.

Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=0, 60/40 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius


Suspect total flow of 60sccm might be too low for the CO2 MFC. Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax. Fitted Tmax was 573, edge is from 560 to 600ns. Garfield calcs give 560ns for 60/40, 573ns (ta-daa!) for 59/41, 588ns for 28/42 and 610ns for 57/43. It's easiest to look at resolution far from the wire to find the best gas mix model.

Resolution for 1975V, drift time > 240ns (~approx 5mm), different gas mix in Garfield, same thresholds and pedestal as before, tz_add=6ns
60/40
59/41
58.5/41.5
58/42
Resolution for 1975V, same thresholds and pedestal as before, tz_add=6ns, 58.5/41.5 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius

Find what to use for local pedestal

Now look at timing. Where to take local pedestal. These are plots of change in ADC value vs time relative to the threshold crossing time. Nothing to do with Garfield or tz_add but already have a good default in place (take pedestal at np=13 minisamples before threshold crossing) so small changes here are a fine tuning.

1975V 59/41 Garfield etc Change in ADC value vs time relative to the threshold crossing time (in units of 8ns/5)
1975V, hit thr 5 ped-sigma
1975V, hit thr 6 ped-sigma

Mean goes through 0.5 and 1 at -16.4 and -13.7 for thr=5sigma, -17.6 and -14.9 for thr=6sigma

1975V 58.5/41.5 Garfield etc Resolution for different local pedestal points, tz_add=6ns
10, 15, and 20 points (1.6ns each) before hit threshold
15, 16, and 17 points (1.6ns each) before hit threshold


1975V 58.5/41.5 Garfield etc Resolution for different local pedestal points, tz_add=7ns
10, 15, and 20 points (1.6ns each) before hit threshold
15, 16, and 17 points (1.6ns each) before hit threshold


Resolution for 1975V, 58.5/41.5 Garfield, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-16, tz_add=7ns, tfix=1000
1975V, resolution ch 20
1975V, resolution vs hit radius

Reassign early hits

Now include previously ignored hits which arrived before tzero+tz_add, assign these to a fixed drift time

1975V 58.5/41.5 Garfield etc Resolution for np=16, tz_add=7ns, early hits assigned to 0ns, 1ns, 2ns
tfix = 0, 1, 2ns


Resolution for 1975V, 58.5/41.5 Garfield, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-16, tz_add=7ns, tfix=1ns
1975V, resolution ch 20
1975V, resolution vs hit radius
Resolution for 1975V, 58.5/41.5 Garfield, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-16, tz_add=7ns, tfix=2ns
1975V, resolution ch 20
1975V, resolution vs hit radius