CDC prototype afterpulses

From GlueXWiki
Jump to: navigation, search

Cosmics 50/50 Ar/CO2

Look for afterpulses by plotting ADC values with samples aligned so that the threshold-crossing is in the same bin for all events. If present they would be approx 100 samples after the start of the first pulse, with Fe55 these started to appear at around 2050V, see Event_pics, but nothing is apparent with cosmics at 2100V.

Events at 2100V
Events at 2100V (zoom in on pedestal)

From an old and short 2200V run

Events at 2200V
Events at 2200V (zoom in on pedestal)