PMT rate tests
PMT Rate Tests
We work here with two PMT assemblies H10534MOD. With the second assembly having a modified base with an additional pre-amplifier powered by the applied HV.
Test with Reference PMT
We test the signal output of the Hamamatsu PMT assembly H10534MOD. This is based on a 10 stage tube with two outputs. The test are performed using a reference PMT assembly in a laser beam where the light output of the laser is adjusted to generate a signal in the reference PMT assembly that is similar to what is seen by the PMTs in the TOF paddle at 1550V operating voltage. One can clearly see that at 5MHz the signal is distorted as the PMT can not handle the increased current. At this point the current in the PMT increased by about 3uA from initially 253uA at 1kHz. At 2MHz the current was still at 253uA.
- Reference PMT with laser output at 6.12
Test with Modified PMT base (preamp)
The modifications to the base are as follows: 2MOhm resistor between photo cathode and Dynode 7, 150kOhm resistor between dynode 7 and dynode 8 and a 420kOhm resistor parallele to the 1MOhm resistor that runs in series with a 2MOhm resistor but all in parallel with the capacitor of the last transitor stage.
- Modified PMT with preamp and with laser output at 6.12
- PMT HV 1300V 1kHz
- PMT HV 1100V 1kHz
- PMT HV 1100V 1kHz zoom
- PMT HV 1100V 1MHz
- PMT HV 1100V 1MHz zoom
- PMT HV 1100V 2MHz
- PMT HV 1100V 2MHz zoom Note the DC offset of +13mV
- PMT HV 1100V 5MHz
- PMT HV 1100V 5MHz zoom Note the DC offset of +23mV
- PMT HV 1100V 10MHz
- PMT HV 1100V 10MHz zoom Note the DC offset of +23mV
- PMT HV 1000V 10MHz
- PMT HV 1000V 10MHz zoom Note DC offset of 18mV
- PMT HV 1000V 100kHz
- PMT HV 1000V 100kHz zoom
- PMT HV 1300V 1kHz