Resolution is calculated as fitted drift distance - measured distance, so it is positive when the measured drift time and distance are too small.
 
Resolution is calculated as fitted drift distance - measured distance, so it is positive when the measured drift time and distance are too small.
 
In this case they are slightly too small at small drift distance but too large at large drift distance. This could be a problem with the gas mix.  For reference, plot res vs distance for tz=5,7,and 0.
 
In this case they are slightly too small at small drift distance but too large at large drift distance. This could be a problem with the gas mix.  For reference, plot res vs distance for tz=5,7,and 0.
 +
 +
 +
{| border="0" cellpadding="2"
 +
|+ Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=8, 60/40 Garfield
 +
|[[Image:run_31842_res20_tz8_6040.png|thumb|x250px|1975V, resolution ch 20]]
 +
|[[Image:run_31842_resx20_tz8_6040.png|thumb|x250px|1975V, resolution vs hit radius]]
 +
|}
 +
Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times.  This would probably have been obvious from the tz=0 data (below).
 +
 +
{| border="0" cellpadding="2"
 +
|+ Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=0, 60/40 Garfield
 +
|[[Image:run_31842_res20_tz0_6040.png|thumb|x250px|1975V, resolution ch 20]]
 +
|[[Image:run_31842_resx20_tz0_6040.png|thumb|x250px|1975V, resolution vs hit radius]]
 +
|}
  
  
 
Suspect total flow of 60sccm might be too low for the CO2 MFC.  Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax.
 
Suspect total flow of 60sccm might be too low for the CO2 MFC.  Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax.

Revision as of 16:54, 11 September 2012

60/40 Ar/CO2 and cosmics, prototype horizontal


1800V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
1900V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
1950V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma
1975V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma
2000V, 60/40 gas, prototype horizontal
Pedestal mean of samples 0-99, ch20, over 1 day
Pedestal std. dev. of samples 0-99, ch20, over 1 day
Max amplitude ch20, hit threshold 4sigma
Max amplitude ch20, hit threshold 5sigma


60/40 gas, prototype horizontal, 4sigma (gray) and 5sigma (blue) hit threshold, fit range 6sigma+
1950V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (64, 79, 110)
1975V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (63, 84, 124)
2000V, lines show 99.7%, 99% and 95% of landau summed 0-4000 (75, 101, 154)
Drift time for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=0
1975V, hit threshold 5sigma, time thresholds 4 and 1 sigma Estimate of t0 is 88 and tmax is 661, difference is 573, straight-line intercept is 81.7, to match histogram fit t0, use tz_add 6.4
Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=6, 60/40 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius

This does not look nearly as good as 50/50 gas mix. Resolution is calculated as fitted drift distance - measured distance, so it is positive when the measured drift time and distance are too small. In this case they are slightly too small at small drift distance but too large at large drift distance. This could be a problem with the gas mix. For reference, plot res vs distance for tz=5,7,and 0.


Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=8, 60/40 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius

Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times. This would probably have been obvious from the tz=0 data (below).

Resolution for 1975V, hit thresholds 5sigma, timing thresholds 4sigma, 1sigma, pedestal at t(t4sig)-13, tz_add=0, 60/40 Garfield
1975V, resolution ch 20
1975V, resolution vs hit radius


Suspect total flow of 60sccm might be too low for the CO2 MFC. Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax.