CDC 19 12 2007
Interposer, preAmp, and signal cable
This is the continuation of CDC_12_12_2007.
Thanks to Fernando we have a new reference measurement; he measured the noise of the interposer+preAmp+signal cable: his measurement is shown in figure 1.
The differential signal was measured with the cable between the CDC-straw-wire and the HV-distribution board disconnected on the CDC side. The measured signal is shown in figure 2.
No oscillations are observed! If one compares to figure 1 the comparison is reasonable, but I think Fernando has a better oscilloscope!
For the sake of completeness, the cable between the CDC and the HV distribution board was also disconnected on the side of the HV-board, this signal can be seen in figure 3.
Again, no oscillations.
Conclusion: the oscillations are caused by the input signals on the preAmp