Difference between revisions of "CDC 60 40"
Line 72: | Line 72: | ||
|[[Image:run_31842_resx20_tz8_6040.png|thumb|x250px|1975V, resolution vs hit radius]] | |[[Image:run_31842_resx20_tz8_6040.png|thumb|x250px|1975V, resolution vs hit radius]] | ||
|} | |} | ||
− | Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times. This would probably have been obvious from the tz=0 data (below) where all the measured drift distances are too large. "Measured" drift distances are actually interpolated from a table of distance calculated for a given gas mix and a range of time measurements. So for this data I should use Garfield tables for a gas mix with | + | Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times. This would probably have been obvious from the tz=0 data (below) where all the measured drift distances are too large. "Measured" drift distances are actually interpolated from a table of distance calculated for a given gas mix and a range of time measurements. So for this data I should use Garfield tables for a gas mix with slower drift velocity, ie less Ar. |
{| border="0" cellpadding="2" | {| border="0" cellpadding="2" | ||
Line 82: | Line 82: | ||
Suspect total flow of 60sccm might be too low for the CO2 MFC. Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax. | Suspect total flow of 60sccm might be too low for the CO2 MFC. Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax. | ||
− | Fitted Tmax was 573, edge is from 560 to 600ns. Garfield calcs give 560ns for 60/40, | + | Fitted Tmax was 573, edge is from 560 to 600ns. Garfield calcs give 560ns for 60/40, 610ns for 57/43. Not sure the fitted Tmax is accurate enough to pick the gas mix given the slope at the high end of drift time histo. |
Revision as of 15:19, 12 September 2012
60/40 Ar/CO2 and cosmics, prototype horizontal
This does not look nearly as good as 50/50 gas mix. Resolution is calculated as fitted drift distance - measured (interpolated) distance, so it is positive when the measured drift time and distance are too small. In this case they are slightly too small at small drift distance but too large at large drift distance. This could be a problem with the gas mix. For reference, plot res vs distance for tz=5,7,and 0.
Increasing tz_add to 8ns shifts the resolution up at earlier times but it hardly moves at later times. This would probably have been obvious from the tz=0 data (below) where all the measured drift distances are too large. "Measured" drift distances are actually interpolated from a table of distance calculated for a given gas mix and a range of time measurements. So for this data I should use Garfield tables for a gas mix with slower drift velocity, ie less Ar.
Suspect total flow of 60sccm might be too low for the CO2 MFC. Increase to 80sccm and repeat. Also try Garfield for different gas mixes and look at tmax.
Fitted Tmax was 573, edge is from 560 to 600ns. Garfield calcs give 560ns for 60/40, 610ns for 57/43. Not sure the fitted Tmax is accurate enough to pick the gas mix given the slope at the high end of drift time histo.