Difference between revisions of "Minutes-3-12-2015"

From GlueXWiki
Jump to: navigation, search
 
(10 intermediate revisions by the same user not shown)
Line 14: Line 14:
 
# CDC update (Mike)
 
# CDC update (Mike)
 
# FDC update [http://www.jlab.org/Hall-D/detector/fdc/spring2015run/HVfdc2_c1_p2.png], [http://www.jlab.org/Hall-D/detector/fdc/corrosion/cathode_from_spare_Package.JPG] (Lubomir)   
 
# FDC update [http://www.jlab.org/Hall-D/detector/fdc/spring2015run/HVfdc2_c1_p2.png], [http://www.jlab.org/Hall-D/detector/fdc/corrosion/cathode_from_spare_Package.JPG] (Lubomir)   
# Electronics (Fernando, Cody, Chris, Nick)
+
# Electronics: logbook entry [https://logbooks.jlab.org/entry/3325323 3325323], [https://logbooks.jlab.org/entry/3325070 3325070] (Fernando, Cody, Chris, Nick)
 
# Engineering (Dave)
 
# Engineering (Dave)
 
# Other
 
# Other
 
<!--
 
  
 
= Minutes =
 
= Minutes =
Line 26: Line 24:
 
== Gas issues ==
 
== Gas issues ==
  
- CO2 was monitored for two weeks. Two sets of measurements are shown in the above plots. First plot: starting with certified gas (60% CO2) in EEL126 then moving the sensor (without disconnecting from the laptop) to the Hall and sensing first FDC1 for several days and at the end switching the solenoid valve to the CDC. Shown are the atmospheric pressure, sensor results and the ones corrected for the pressure: ~(dleta_p/p)^n, where the power n was chosen, on the first plot, to make the calibration measurements flat (n=2.4). Second plot is the same except n=1.4 which makes the CDC measurements plat. Third and fourth plots represent the second set of measurements, starting with 5 days sensing CDC and then few days FDC1.  
+
- CO2 was monitored for another 4 days: the first plot is with a correction function with a power of 2.4, second with 1.4, as explained in the previous meeting. First, the sensor was connected to the CDC for one day(Thursday-Friday), then moved to EEl126. There the bottle was almost empty and as it turned out the CO2 was lower than 60%. On Sunday the bottle was replaced and by Tuesday the values were close to 60%. At that time the fan failed - as the instructions say it has 1000h life and we were using it one month already. We didn't learn a lot from these measurements, except that the change of the gas factor, done a week ago, showed almost no effect (within 0.5-1%, as expected) on the gas mixture. If we want to continue monitoring CO2 we need to buy another pump ($200) that will work for another month, or try to find a better pump. Beni suggested using the pump that Bill made for the O2 measurements.  
  
- Discussions:generally FDC is off by 4-5% consistent with the time-to-distance results. CDC is off by 8-9-10% while the time-to-distance from the data requires ~6% off (or ~44%). Proposal: to calibrate the sensor with a gas that is closer to the 40% CO2 as measured in the CDC. Need more calibration measurements to be sure the pressure correction is correct. If assume n=1.4 CDC is more stable then the FDC, could be due to the higher flow.  
+
- Beni did different changes in the settings of the mass flow controllers used for the CDC mixing and was looking at the maximum drift time. Changing the gas factor didn't show any difference (within 1%). Then he changed the controller mode from "mixing" to independent flow control, also made the flow rates in and out of the mixing tank very similar. With the latest modification Beni found a slight change in the drift time but in the opposite direction, i.e. shorter. Next steps: change the gas factors to 1 to see if it will make any difference. Beni suspects something is wrong with the controller boxes.    
  
- Gas gain calibrations: using Fe source measurements were done in the EEL126 with the spare package at three different HV (first plot). They will be compared to the results from the Hall with the source put at the downstream side of FDC. Second plot shows some change of the gas gain with the atmospheric pressure. Since the gas properties scale with field/pressure, it goes in the right direction.
+
- Lubomir did measurements in the Hall with an iron source attached at the downstream side of the FDC. There were two sets of measurements (Friday and Monday) at different atmospheric pressures, both sets at three HV values 2100-2150-2200V. Results will be presented on the next meeting, but we do see indeed higher amplitudes in the Hall, compared to the measurements with the spare package, consistent with the lower CO2 percentage.  
  
== CDC update ==
+
== FDC update ==
  
- CDC was left without alcohol for a week - very small change (~1%) in the drift-time spectrum.
+
- The chiller is working fine so far. There was a problem with one HV channel (FDC2, Cell1) that tripped on Wednesday night and then it took three days to condition it and set it back. In the plot attached you see that on Wednesday the humidity was >55%, which combined with the water on the stairs from the melting snow and the fact that the AC was not working at that time, may have caused the problem. Fortunately, next three days the humidity was as low as 20%.  
  
== FDC update ==
+
- Sean was interested in the progress with the FDC alignment. Lubomir: the procedure is first, to align all the elements withing the package: 8 parameters per cell or 48 per package. Three packages have been aligned already and it will take one more week to align the last one. It turned out that the procedure converges only if there's enough statistics for each iteration, at least a quoter of the total, that's why it takes time. The last step will be to align the packages between each other, but there are only three parameters per package, so it will work faster.
  
- The Chiller came back from the repair and was connected to the FDC, so far works fine.
+
- Today we did (together with Dave and Casey) a small modification in the spare package. We used the opportunity to look at the cathode: as seen from the picture above, there are no hints for corrosion in the place where the O-ring goes over the connector traces. The spare package was sitting in EEL126 for two years.
 
+
- One HV channel tripped, took several days to set it back.
+
  
 
== Electronics ==
 
== Electronics ==
  
- Beni: new f125 firmware on all modules. CDC was read out for 24 hours at 30kHz without crashing.  
+
- Beni checked the fADC crates with pulser (Mike's code) and found ~10 bad channels in the FDC. 6 of them had some connector issues and were fixed by Chris and Nick, 4 are still bad (first log entry).
 
+
== Engineering ==
+
  
- Dave: CDC flow sensor connected and operational. FDC chiller shut off also connected and tested.
+
- Lubomir looked at the data (first runs after the chiller was installed) and made a list of bad channels (second log entry linked above). There's nothing new or unexpected, these are 0.6% of the total channel number. Noticeably, there are 8 new bad wire channels sitting on one ASIC in the pre-amp. Fernando wanted to have a code, similar to the one for the fADCs, to check the TDCs with a pulser. Bryan has implemented already this feature in his libraries.  
  
-->
+
- Beni: one FDC crate (rocfdc9) is crushing from time to time, he suspects broken SD.

Latest revision as of 18:34, 12 March 2015

March 12, 2015 FDC+CDC meeting

Connection

  1. Instructions for Bluejeans meeting connection
  2. FDC meeting ID: 290664653
  3. To join via a Web Browser, go to the page [1] https://bluejeans.com/290664653.

Agenda

  1. Gas issues
    • CO2 monitoring [2],[3]
    • FDC gain measurements
  2. CDC update (Mike)
  3. FDC update [4], [5] (Lubomir)
  4. Electronics: logbook entry 3325323, 3325070 (Fernando, Cody, Chris, Nick)
  5. Engineering (Dave)
  6. Other

Minutes

Participants: Curtis, Naomi (CMU), Sean (NU), Fernando, Dave, Chris, Nick, Simon, Beni, and Lubomir (JLab).

Gas issues

- CO2 was monitored for another 4 days: the first plot is with a correction function with a power of 2.4, second with 1.4, as explained in the previous meeting. First, the sensor was connected to the CDC for one day(Thursday-Friday), then moved to EEl126. There the bottle was almost empty and as it turned out the CO2 was lower than 60%. On Sunday the bottle was replaced and by Tuesday the values were close to 60%. At that time the fan failed - as the instructions say it has 1000h life and we were using it one month already. We didn't learn a lot from these measurements, except that the change of the gas factor, done a week ago, showed almost no effect (within 0.5-1%, as expected) on the gas mixture. If we want to continue monitoring CO2 we need to buy another pump ($200) that will work for another month, or try to find a better pump. Beni suggested using the pump that Bill made for the O2 measurements.

- Beni did different changes in the settings of the mass flow controllers used for the CDC mixing and was looking at the maximum drift time. Changing the gas factor didn't show any difference (within 1%). Then he changed the controller mode from "mixing" to independent flow control, also made the flow rates in and out of the mixing tank very similar. With the latest modification Beni found a slight change in the drift time but in the opposite direction, i.e. shorter. Next steps: change the gas factors to 1 to see if it will make any difference. Beni suspects something is wrong with the controller boxes.

- Lubomir did measurements in the Hall with an iron source attached at the downstream side of the FDC. There were two sets of measurements (Friday and Monday) at different atmospheric pressures, both sets at three HV values 2100-2150-2200V. Results will be presented on the next meeting, but we do see indeed higher amplitudes in the Hall, compared to the measurements with the spare package, consistent with the lower CO2 percentage.

FDC update

- The chiller is working fine so far. There was a problem with one HV channel (FDC2, Cell1) that tripped on Wednesday night and then it took three days to condition it and set it back. In the plot attached you see that on Wednesday the humidity was >55%, which combined with the water on the stairs from the melting snow and the fact that the AC was not working at that time, may have caused the problem. Fortunately, next three days the humidity was as low as 20%.

- Sean was interested in the progress with the FDC alignment. Lubomir: the procedure is first, to align all the elements withing the package: 8 parameters per cell or 48 per package. Three packages have been aligned already and it will take one more week to align the last one. It turned out that the procedure converges only if there's enough statistics for each iteration, at least a quoter of the total, that's why it takes time. The last step will be to align the packages between each other, but there are only three parameters per package, so it will work faster.

- Today we did (together with Dave and Casey) a small modification in the spare package. We used the opportunity to look at the cathode: as seen from the picture above, there are no hints for corrosion in the place where the O-ring goes over the connector traces. The spare package was sitting in EEL126 for two years.

Electronics

- Beni checked the fADC crates with pulser (Mike's code) and found ~10 bad channels in the FDC. 6 of them had some connector issues and were fixed by Chris and Nick, 4 are still bad (first log entry).

- Lubomir looked at the data (first runs after the chiller was installed) and made a list of bad channels (second log entry linked above). There's nothing new or unexpected, these are 0.6% of the total channel number. Noticeably, there are 8 new bad wire channels sitting on one ASIC in the pre-amp. Fernando wanted to have a code, similar to the one for the fADCs, to check the TDCs with a pulser. Bryan has implemented already this feature in his libraries.

- Beni: one FDC crate (rocfdc9) is crushing from time to time, he suspects broken SD.