Monitoring System Requirements

From GlueXWiki
Revision as of 17:01, 8 December 2010 by Zisis (Talk | contribs)

Jump to: navigation, search

Monitoring Requirements Table

DRAFT Table of Parameters and Requirements
Characteristic Fcal Bcal TOF Start
Light uniformity channel-to-channel ≤ 15%
Dynamic range of input (number of photoelectrons) typically 100 100 up to 800
Requirement on absolute light input none (PRIMEX might have) 150 pe on module
Variation in arrival times of cal pulses ~3 ns ≤ 2 ns ~3ns ~3ns
Time stability gain and timing (one channel) ≤ 10% in 6 months
Temperature stability (likely cause of time stability) 1o causes 3% gain change
Location of calibration input From front to monitor rad damage light guide tapered side
Geometrical constraints Approximately 90% of front 4x4 cm2 area available packing of WC and LGs, LED fitting, cable routing
  1. Comment: Time (and temperature) stability most important.
  2. Comment: Option of delivering a range of light intensities important (e.g. using various filter settings).
  3. Comment from Werner: The flexibility in the use of the LED would be very useful for the start counter. Christina: Still need to test stability.
  4. Questions to Matt: What uniformity could one expect from the plexiglass input system? How many fibers were used to pulse the plexiglass, i.e. how often were they spaced?
  5. Comment: Matt drew attention to an LED pulser by Paul Smith which might be a useful reference.