Difference between revisions of "JEF meeting, January 13, 2022"

From GlueXWiki
Jump to: navigation, search
(Agenda)
 
Line 34: Line 34:
 
**[https://userweb.jlab.org/~aasatury/Fcal_LMS/beam_test2/ Beam test]
 
**[https://userweb.jlab.org/~aasatury/Fcal_LMS/beam_test2/ Beam test]
 
* '''Simulations/Analysis'''
 
* '''Simulations/Analysis'''
 +
 +
=Minutes=
 +
Attendees: Bill, Arshak, Sasha, Igal, Zisis, Liping, Jane, Olga, Jon, Eugene, Malte, Axel, S.T.
 +
* We have received all 835 Crytur crystals we expected to receive last year
 +
** All wrapped
 +
** 95 more expected in May
 +
* Grand total number of fabricated modules = 1408 (only 3 more crystals available for fabrication in lab)
 +
* COMCAL removed from hall
 +
** Modules need to be at least partially disassembled
 +
*** Replace 25 micron brass straps with 50 micron brass straps? re-wrap?
 +
*** Some tedlar had grease contamination due use of grease for optical coupling
 +
*** PMT shielding: replace FR4 with soft iron
 +
*** Need to remove plastic cap holding optical fiber
 +
* Plan to install 5x5 prototype using COMCAL crystals in hall
 +
** Instrument with new electronics to test hight rate response
 +
** Check 3x amplifier in new PMT base design
 +
*** Have 25 bases in house
 +
*** Test noise level
 +
* Plans for module installation coming together:  student help, procedures, shims and alignment tools, ...
 +
** Expect some help from imaging group
 +
* Arshak continuing to work on LMS beam test using 3x3 matrix
 +
** Plexiglas plate in front with similar design to actual plate but smaller size
 +
** Middle fully instrumented with FCAL-2 module, sides blanks with varying light access in the front (close/open faces)
 +
** Measured cross talk using PS setup selecting particular PS tiles
 +
** Cross talk level <~0.5% integrating over all amplitude, but most of the amplitude is near pedestal
 +
*** Suggest look at amplitude dependence
 +
*** MC simulation of effect?
 +
* Also discussed using individual fibers for LMS
 +
** Not enough light from 250 micron fibers?  Thicker diameter fibers more difficult to bend
 +
** Light uniformity?

Latest revision as of 09:01, 17 January 2023

Meeting time and location:

  • 9:30 am EST (JLab time) remote only
  • Zoom meeting ID: 161 050 9142 Passcode: 004965

Agenda

Minutes

Attendees: Bill, Arshak, Sasha, Igal, Zisis, Liping, Jane, Olga, Jon, Eugene, Malte, Axel, S.T.

  • We have received all 835 Crytur crystals we expected to receive last year
    • All wrapped
    • 95 more expected in May
  • Grand total number of fabricated modules = 1408 (only 3 more crystals available for fabrication in lab)
  • COMCAL removed from hall
    • Modules need to be at least partially disassembled
      • Replace 25 micron brass straps with 50 micron brass straps? re-wrap?
      • Some tedlar had grease contamination due use of grease for optical coupling
      • PMT shielding: replace FR4 with soft iron
      • Need to remove plastic cap holding optical fiber
  • Plan to install 5x5 prototype using COMCAL crystals in hall
    • Instrument with new electronics to test hight rate response
    • Check 3x amplifier in new PMT base design
      • Have 25 bases in house
      • Test noise level
  • Plans for module installation coming together: student help, procedures, shims and alignment tools, ...
    • Expect some help from imaging group
  • Arshak continuing to work on LMS beam test using 3x3 matrix
    • Plexiglas plate in front with similar design to actual plate but smaller size
    • Middle fully instrumented with FCAL-2 module, sides blanks with varying light access in the front (close/open faces)
    • Measured cross talk using PS setup selecting particular PS tiles
    • Cross talk level <~0.5% integrating over all amplitude, but most of the amplitude is near pedestal
      • Suggest look at amplitude dependence
      • MC simulation of effect?
  • Also discussed using individual fibers for LMS
    • Not enough light from 250 micron fibers? Thicker diameter fibers more difficult to bend
    • Light uniformity?